23

22

Radiated RF
IEC
61000-4-3

3V/m
80kHz to
2.5GHz

3V/m

d

=1.2

P

80MHz to 800MHz

d

=2.3

P

800MHz

to 2.5MHz

Here

P

is the maximum output power rating

of the transmitter in watts (W) according to
the transmitter manufacturer and

d

is the

recommended separation distance in meters

(m). Field strengths form fixed RF transmitters,

as determined by an electromagnetic site
survey, should be less than the compliance

b level in each frequency range. Interference

may occur in the vicinity of equipment
marked with the following symbol:

NOTE 1 At 90MHz and 800MHz, the higher frequency range applies.
NOTE 2 These guidelines may not apply in all situations.
Electromagnetic propagation is affected by absorption and reflection
from structures, objects and people.
a
Field strengths from fixed transmitters, such as base stations for radio
(cellular/cordless)

telephones and land mobile radios, amateur radio,

AM and FM radio broadcast and TV broadcast cannot be predicted
theoretically with accuracy. To assess the electromagnetic
environment due to fixed RF transmitters, an electromagnetic site
survey should be considered. If the measured field strength in the
location in which the thermometer is used exceeds the applicable RF
compliance level above, the thermometer should be observed to verify
normal operation. If abnormal performance is observed, additional
measures may be necessary, such as reorienting or relocating the
thermometer.
b
Over the frequency range 150kHz to 80MHz, field strengths should be
less than 3V/m.

Recommended separation distances between portable and mobile RF
communications equipment and the infrared thermometer

The infrared thermometer is intended for use in an electromagnetic
environment in which radiated RF disturbances are controlled.
The customer or the user of the infrared thermometer can help prevent
electromagnetic interference by maintaining a minimum
distance between portable and mobile RF communications

equipment

(transmitters) and the infrared thermometer as recommended below,
according to the maximum output power of the communications
equipment.

Rated maximum
output power of
transmitter
W

Separation distance according to frequency of
transmitter
m

Rated
maximum
output power
of transmitter
W

150kHz to
80MHz

d

=1.2

P

80MHz to
800MHz

d

=1.2

P

800MHz to
2.5GHz

d

=2.3

P

0.01

0.01

0.12

0.23

0.1

0.1

0.38

0.73

1

1

1.2

2.3

10

10

3.8

7.3

100

100

12

23

For transmitters rated at a maximum output power not listed above, the recommended 
separation distance d in meters (m) can be estimated using the equation applicable to 
the frequency of the transmitter, where P is the maximum output power rating of the 
transmitter in watts (W) according to the transmitter manufacturer.

Summary of Contents for 607

Page 1: ... Trust ensures that our technically efficient team works dynamically and tirelesslytoprovidethebestofthemedicaldevicestoourclients Theproductsthatwehave toofferaresuitablydesignedforuseathomes laboratoriesandhospitals Our ground breaking solutions allow you to monitor your health in the easiest ways possible In today s era when all of our lives are too hassled to handle it becomes a bit difficult topay...

Page 2: ...y Replacement 9 ºC ºF Conversion 3 Room object Temperature Measurement 11 Turn off Thermometer 1 WARNINGS AND PRECAUTIONS 5 DISPLAY DESCRIPTION 4 Indications for use 3 Operating principle 2 Structure 1 Overview 7 TIPS TO MAKE CORRECT TEMPERATURE MEASUREMENTS 8 MULTI MEASUREMENTS REQUIREMENTS 9 KEY FACTORS THAT INFLUENCE EAR TEMPERATURE READINGS 10 CARE AND MAINTENANCE 11 SPECIFICATIONS 13 SYMBOLS 1...

Page 3: ...ildren to take their temperature unattended When You Need to Receive Medical Attention 1 Please consult your doctor if you see symptoms such as unexplained irritability frail have a weakened immune system or neonates and infants vomiting diarrhea dehydration changes in appetite or activity seizure muscle pain shivering stiff neck pain when urinating etc 2 People who are on antibiotics analgesics or...

Page 4: ... ages and the eardrum mode isindicatedforpeopleabovethreemonths Note Do not use the thermometer if the ear is infected with otitis or suppuration 3 FEATURES Ÿ 1 second results Takes less than 1 second for temperature measurement Ÿ Intelligent Light Indicator Normal body temperature shows green low fever shows orange and high fever shows red Ÿ Sound and mute function Do not worry about the prompt b...

Page 5: ...appears please replace the battery Note 6 HOW TO USE YOUR THERMOMETER The Dr Trust Handy Forehead and Ear Thermometer 607 is lightweight gentle and easy to use device It comes with all the things needed to get a thorough reading When using the thermometer for the rst time please check the batteries and press the Measure button to check it is ready for use Ÿ Release the bu on when a beep is heard 1...

Page 6: ...al is straight so that the temperature probe can receive an infrared ray from the eardrum ü Be careful when taking temperature on a child whose ear canal is small ü Children aged 1 year to adult Pull the ear up and back ü Do not force the thermometer into the ear canal Otherwise the ear canal may get injured Note Ÿ Insert the probe straight into the ear canal Ÿ Please make sure that the ear is cle...

Page 7: ...ORTANT It is dangerous to make a self diagnosis or self treatment based on the obtained measurement results For such purposes please consult a doctor Temperature Reading In forehead or ear mode T indicates a temperature reading Intelligent Light Indicator one long beep 1 If 32 C T 37 3 C 89 6 F T 99 2 F the green light will last for 3 seconds with fever 3 short beeps and the value in LCD flickers w...

Page 8: ...tions on the forehead Ÿ Holding the thermometer for too long in the hand before taking a measurement can cause the device to warm up This means the measurement could be incorrect Ÿ Patients and the thermometer should stay in steady state room condition for at least 30 minutes Ÿ A child s normal temperature can be as high as 99 9 F 37 7 or as low as 97 0 F 36 11 Please note that this unit reads 0 5...

Page 9: ...r use abrasive cleaning agents thinners or benzene for cleaning and never immerse the instrument in water or other cleaning liquids Ÿ Take care not to scratch the surface of the LCD screen Ÿ Ensure that no liquid enters the interior of the thermometer 12 ERROR AND TROUBLESHOOTING Symptom Possible Cause Description Solution Failed to power on The battery level is too low Replace with a new battery ...

Page 10: ... solid foreign objects of 12 5 mm Ø and greater 2 If keep the thermometer in 15 degree angle it still can prevent the water drop 14 EMC INFORMATION Guidance and manufacturer s declaration electromagnetic emissions The infrared thermometer is intended for using in the electromagnetic environment specified below The customer or the user of the infrared thermometer should assure that it is used in suc...

Page 11: ...5 cycles 70 UT N A Mains power quality should be that of a typical commercial or hospital environment If the user of the infrared thermometer requires continued operation during power mains Guidance and manufacturer s declaration electromagnetic immunity The infrared thermometer is intended for use in the electromagnetic environment specified below The customer or the user of the infrared thermomet...

Page 12: ...er should be observed to verify normal operation If abnormal performance is observed additional measures may be necessary such as reorienting or relocating the thermometer b Over the frequency range 150kHz to 80MHz field strengths should be less than 3V m Recommended separation distances between portable and mobile RF communications equipment and the infrared thermometer The infrared thermometer is...

Page 13: ...ts and people 15 WARRANTY AND AFTER SALE SERVICE 2 Failure resulting from an unexpected dropping during application or transportation The batteries the packaging and any damage caused by improper use are not covered by the warranty Excluding the following user caused failures 1 Failure resulting from unauthorized disassembly and modi cation 3 Failure resulting from not following the instructions i...

Reviews: