Measurements

Forehead temperature

36.1°C to 37.5°C (97°F to 99.5°F)

Ear temperature

35.8°C to 38°C (96.4°F to 100.4°F)

Oral temperature

35.5°C to 37.5°C (95.9°F to 99.5°F)

Rectal temperature

36.6°C to 38°C (97.9°F to 100.4°F)

Axillary temperature

34.7°C to 37.3°C (94.5°F to 99.1°F)

2) Structure

The thermometer consists of a shell, an LCD, an operation button, an infrared 
temperature sensor, a microprocessor and a beeper.

3) Operating principle

The infrared temperature sensor collects infrared energy emitted by the eardrum or the 
skin surface. The collected energy is converted into a temperature reading by the 
thermopiles and measurement circuits.

4) Indications for use

The Dual-mode Infrared Thermometer is intended for the measurement of human body 
temperatures. The forehead mode is indicated for people of all ages and the eardrum mode 
is indicated for people above three months.

Note

Do not use the thermometer if the ear is infected with otitis or suppuration.

3. FEATURES

Ÿ

1 second results: Takes less than 1 second for temperature measurement

Ÿ

Intelligent Light Indicator: Normal body temperature shows green, low fever shows 
orange and high fever shows red

Ÿ

Sound and mute function: Do not worry about the prompt beep sound that may 
disturb your baby. It can be used while a child is asleep. 

Ÿ

Multi-functional: Can measure ear, forehead, room, milk, water and object 
temperature.

Ÿ

Accurate and reliable

Ÿ

35 sets of memories

Ÿ

Easy to recall historical measurements data 

Ÿ

Fever alarm function displayed in orange and red light

Ÿ

Easy switching between mute and un-mute mode

Ÿ

Easy switching between ºC and ºF

Ÿ

Auto shut-down after 30 seconds and power-saving

Ÿ

Completely safe with passive infrared receiving technology

Ÿ

Adopt the latest algorithm, temperature compensation function

Ÿ

Zero emission of radiation during usage     

Summary of Contents for 607

Page 1: ... Trust ensures that our technically efficient team works dynamically and tirelesslytoprovidethebestofthemedicaldevicestoourclients Theproductsthatwehave toofferaresuitablydesignedforuseathomes laboratoriesandhospitals Our ground breaking solutions allow you to monitor your health in the easiest ways possible In today s era when all of our lives are too hassled to handle it becomes a bit difficult topay...

Page 2: ...y Replacement 9 ºC ºF Conversion 3 Room object Temperature Measurement 11 Turn off Thermometer 1 WARNINGS AND PRECAUTIONS 5 DISPLAY DESCRIPTION 4 Indications for use 3 Operating principle 2 Structure 1 Overview 7 TIPS TO MAKE CORRECT TEMPERATURE MEASUREMENTS 8 MULTI MEASUREMENTS REQUIREMENTS 9 KEY FACTORS THAT INFLUENCE EAR TEMPERATURE READINGS 10 CARE AND MAINTENANCE 11 SPECIFICATIONS 13 SYMBOLS 1...

Page 3: ...ildren to take their temperature unattended When You Need to Receive Medical Attention 1 Please consult your doctor if you see symptoms such as unexplained irritability frail have a weakened immune system or neonates and infants vomiting diarrhea dehydration changes in appetite or activity seizure muscle pain shivering stiff neck pain when urinating etc 2 People who are on antibiotics analgesics or...

Page 4: ... ages and the eardrum mode isindicatedforpeopleabovethreemonths Note Do not use the thermometer if the ear is infected with otitis or suppuration 3 FEATURES Ÿ 1 second results Takes less than 1 second for temperature measurement Ÿ Intelligent Light Indicator Normal body temperature shows green low fever shows orange and high fever shows red Ÿ Sound and mute function Do not worry about the prompt b...

Page 5: ...appears please replace the battery Note 6 HOW TO USE YOUR THERMOMETER The Dr Trust Handy Forehead and Ear Thermometer 607 is lightweight gentle and easy to use device It comes with all the things needed to get a thorough reading When using the thermometer for the rst time please check the batteries and press the Measure button to check it is ready for use Ÿ Release the bu on when a beep is heard 1...

Page 6: ...al is straight so that the temperature probe can receive an infrared ray from the eardrum ü Be careful when taking temperature on a child whose ear canal is small ü Children aged 1 year to adult Pull the ear up and back ü Do not force the thermometer into the ear canal Otherwise the ear canal may get injured Note Ÿ Insert the probe straight into the ear canal Ÿ Please make sure that the ear is cle...

Page 7: ...ORTANT It is dangerous to make a self diagnosis or self treatment based on the obtained measurement results For such purposes please consult a doctor Temperature Reading In forehead or ear mode T indicates a temperature reading Intelligent Light Indicator one long beep 1 If 32 C T 37 3 C 89 6 F T 99 2 F the green light will last for 3 seconds with fever 3 short beeps and the value in LCD flickers w...

Page 8: ...tions on the forehead Ÿ Holding the thermometer for too long in the hand before taking a measurement can cause the device to warm up This means the measurement could be incorrect Ÿ Patients and the thermometer should stay in steady state room condition for at least 30 minutes Ÿ A child s normal temperature can be as high as 99 9 F 37 7 or as low as 97 0 F 36 11 Please note that this unit reads 0 5...

Page 9: ...r use abrasive cleaning agents thinners or benzene for cleaning and never immerse the instrument in water or other cleaning liquids Ÿ Take care not to scratch the surface of the LCD screen Ÿ Ensure that no liquid enters the interior of the thermometer 12 ERROR AND TROUBLESHOOTING Symptom Possible Cause Description Solution Failed to power on The battery level is too low Replace with a new battery ...

Page 10: ... solid foreign objects of 12 5 mm Ø and greater 2 If keep the thermometer in 15 degree angle it still can prevent the water drop 14 EMC INFORMATION Guidance and manufacturer s declaration electromagnetic emissions The infrared thermometer is intended for using in the electromagnetic environment specified below The customer or the user of the infrared thermometer should assure that it is used in suc...

Page 11: ...5 cycles 70 UT N A Mains power quality should be that of a typical commercial or hospital environment If the user of the infrared thermometer requires continued operation during power mains Guidance and manufacturer s declaration electromagnetic immunity The infrared thermometer is intended for use in the electromagnetic environment specified below The customer or the user of the infrared thermomet...

Page 12: ...er should be observed to verify normal operation If abnormal performance is observed additional measures may be necessary such as reorienting or relocating the thermometer b Over the frequency range 150kHz to 80MHz field strengths should be less than 3V m Recommended separation distances between portable and mobile RF communications equipment and the infrared thermometer The infrared thermometer is...

Page 13: ...ts and people 15 WARRANTY AND AFTER SALE SERVICE 2 Failure resulting from an unexpected dropping during application or transportation The batteries the packaging and any damage caused by improper use are not covered by the warranty Excluding the following user caused failures 1 Failure resulting from unauthorized disassembly and modi cation 3 Failure resulting from not following the instructions i...