With the unit powered off, depress the On/Off after depressing
and holding the Select key to enter the product probe calibration
menu.
ii.
Using the select key, scroll through each probe as indicated
by Select LED indicators.
iii.
Use the up down arrows to add or subtract from the displayed
probe reading. The displayed reading will reflect the raw reading
plus the offset.
iv
. Exit menu by depressing Start/Stop.
v.
Note:
1. Missing probes will be displayed as “OPEn”. 2.
Shorted probes or wires will be displayed as “SHrt” 3.
Recalibration will be necessary if probes are interchanged or
replaced.