IR9-PEN Infrared Thermometer 

Operating Instructions 

Congratulate on your selection of this IR9-PEN thermometer. The 

IR9-PEN thermometer is a non-contact infrared thermometer with 

circle laser.   

Top view 

Side view 

Operation:   

1.

Press ‘Meas.’ Button, directly to power IR9-PEN on and start 

measurement.   

2.

Release finger to hold the last reading and see the last, 

Maximum and Minimum readings every 2 seconds in turn.   

Hold        Maximum          Minimum 

o

C /

 o

F Measurement Unit Conversion 

Take off battery cover and use a sticker to shift the scale switch     

LCD ERROR MESSAGES

IR9-PEN incorporates visual diagnostic messages as follows: 

‘Hi’  or  ’Lo’  is  displayed  when  the 

‘Hi’  when  higher  than  +500

  o

  o

F

and  ‘Lo’  when lower than 

 -33

 o

C (-27

 o

F).

‘Er2’ is displayed when the IR9-PEN is exposed to 

rapid changes in the ambient temperature.   

‘Er1’ is displayed when the ambient temperature of 

the IR9-PEN exceeds –10

 o

C (14

o

F) OR +50

 o

C

(122

 o

F). In both cases you should allow plenty of 

time (minimum 30 minutes ) for the IR9-PEN to 

stabilize to the working/room temperature 0 to +40

o

C (32 to 104

 o

F)

.

For all other error messages it is necessary to reset 

the IR9-PEN. To reset the IR9-PEN, turn the 

instrument off, remove the battery and wait for a 

minimum of one minute, reinsert the battery and 

turn on. If the error message remains please contact 

the Service Department  for further assistance. 

BATTERIES 

The IR9-PEN incorporates visual low battery indication as follows: 

‘Battery OK’: 

measurements 

  are possible 

‘Battery Low’: battery 

needs to be replaced, 

measurements are 

possible

‘Battery Exhausted’:

measurements are 

not possible 

BATTERY REPLACEMENT 

When the ‘Low Battery’ icon indicates the battery is low, the  

battery should  be  replaced  immediately  with  a  AA  battery,

The battery is located under the battery cover at the rear of

. Slide the battery cover out     

. Take  off  the  old  battery  and  put  a  new  AA  battery  into 

battery compartment. 

  WARNING

. The Temp function is not inte nded for taking measurements 

that require for Live & Die decision or Safety related 

applications.

. The temperature reading of IR9-PEN indicates only the 

surface temperature, and not the actual temperature of an 

object. 

. Keep the unit dry and away fr om any liquids and direct 

sunlight or a very hot object. (Getting too close to a very hot 

object will creates the risk of burn injury, please be careful.) 

. Always make sure the probe lens is clean without any damage. 
. Dispose of used battery promptly and keep away from 

children. 

STORAGE & CLEANING 

SPECIFICATION 

Range 

-33 to +500

o

C(-27 to +932

 o

F)

Resolution 

199.9 to -9.9 

o

C /

 o

F: 0.1 

o

C /

 o

F

 Above 200 or below-10 

o

C /

 o

F F: 1 

o

C /

 o

F

Response Time 

1 second 

Infrared Accuracy  

+

2.0% of reading or +

2.0

 o

C which one is 

greater 

Field of View (D:S)  

   

Emissivity    

0.95 fixed 

Battery Life   

30 hours continuous use 

Operation temp. and 

humidity  

 0 to 50

 o

C; < 90RH%        

Storage temp. and 

humidity  

 10 to 40

 o

C ; < 65 RH% 

Battery     

AA   

Display   

LCD 

Dimensions  

150l(L) x 25(W) x 27(H) mm 

Weight   

74grams including battery 

GUARANTED 

The IR9-PEN is guaranteed for a period of one year from the date 

of  purchase  against  mechanical  and  electrical  manufacturing 

defects. There  are  no  user  serviceable  parts  inside  the   

instrument. Any attempted repair by unauthorized  persons voids 

o

 o

F  

C

(932 )

the IR9-PEN.   

9:1 optics ratio

the warranty.

Always keep the reflect mirror and lens clean. 

 

If it is dirty, use a wet soft cloth or wet cotton

swab to clean it softly and make the lens is 

 fully dry before using the IR9 - PEN .

 

 

IR9 -PEN is not waterproof, do NOT 

 

 submerge any part of it in water.

 

 

temperature being measured is

outside of the range of the instrument,

IR9-PEN Infrared Thermometer 

Operating Instructions 

Congratulate on your selection of this IR9-PEN thermometer. The 

IR9-PEN thermometer is a non-contact infrared thermometer with 

circle laser.   

Top view 

Side view 

Operation:   

1.

Press ‘Meas.’ Button, directly to power IR9-PEN on and start 

measurement.   

2.

Release finger to hold the last reading and see the last, 

Maximum and Minimum readings every 2 seconds in turn.   

Hold        Maximum          Minimum 

o

C /

 o

F Measurement Unit Conversion 

Take off battery cover and use a sticker to shift the scale switch     

LCD ERROR MESSAGES

IR9-PEN incorporates visual diagnostic messages as follows: 

‘Hi’  or  ’Lo’  is  displayed  when  the 

‘Hi’  when  higher  than  +500

  o

  o

F

and  ‘Lo’  when lower than 

 -33

 o

C (-27

 o

F).

‘Er2’ is displayed when the IR9-PEN is exposed to 

rapid changes in the ambient temperature.   

‘Er1’ is displayed when the ambient temperature of 

the IR9-PEN exceeds –10

 o

C (14

o

F) OR +50

 o

C

(122

 o

F). In both cases you should allow plenty of 

time (minimum 30 minutes ) for the IR9-PEN to 

stabilize to the working/room temperature 0 to +40

o

C (32 to 104

 o

F)

.

For all other error messages it is necessary to reset 

the IR9-PEN. To reset the IR9-PEN, turn the 

instrument off, remove the battery and wait for a 

minimum of one minute, reinsert the battery and 

turn on. If the error message remains please contact 

the Service Department  for further assistance. 

BATTERIES 

The IR9-PEN incorporates visual low battery indication as follows: 

‘Battery OK’: 

measurements 

  are possible 

‘Battery Low’: battery 

needs to be replaced, 

measurements are 

possible

‘Battery Exhausted’:

measurements are 

not possible 

BATTERY REPLACEMENT 

When the ‘Low Battery’ icon indicates the battery is low, the  

battery should  be  replaced  immediately  with  a  AA  battery,

The battery is located under the battery cover at the rear of

. Slide the battery cover out     

. Take  off  the  old  battery  and  put  a  new  AA  battery  into 

battery compartment. 

  WARNING

. The Temp function is not inte nded for taking measurements 

that require for Live & Die decision or Safety related 

applications.

. The temperature reading of IR9-PEN indicates only the 

surface temperature, and not the actual temperature of an 

object. 

. Keep the unit dry and away fr om any liquids and direct 

sunlight or a very hot object. (Getting too close to a very hot 

object will creates the risk of burn injury, please be careful.) 

. Always make sure the probe lens is clean without any damage. 
. Dispose of used battery promptly and keep away from 

children. 

STORAGE & CLEANING 

SPECIFICATION 

Range 

-33 to +500

o

C(-27 to +932

 o

F)

Resolution 

199.9 to -9.9 

o

C /

 o

F: 0.1 

o

C /

 o

F

 Above 200 or below-10 

o

C /

 o

F F: 1 

o

C /

 o

F

Response Time 

1 second 

Infrared Accuracy  

+

2.0% of reading or +

2.0

 o

C which one is 

greater 

Field of View (D:S)  

   

Emissivity    

0.95 fixed 

Battery Life   

30 hours continuous use 

Operation temp. and 

humidity  

 0 to 50

 o

C; < 90RH%        

Storage temp. and 

humidity  

 10 to 40

 o

C ; < 65 RH% 

Battery     

AA   

Display   

LCD 

Dimensions  

150l(L) x 25(W) x 27(H) mm 

Weight   

74grams including battery 

GUARANTED 

The IR9-PEN is guaranteed for a period of one year from the date 

of  purchase  against  mechanical  and  electrical  manufacturing 

defects. There  are  no  user  serviceable  parts  inside  the   

instrument. Any attempted repair by unauthorized  persons voids 

o

 o

F  

C

(932 )

the IR9-PEN.   

9:1 optics ratio

the warranty.

Always keep the reflect mirror and lens clean. 

 

If it is dirty, use a wet soft cloth or wet cotton

swab to clean it softly and make the lens is 

 fully dry before using the IR9 - PEN .

 

 

IR9 -PEN is not waterproof, do NOT 

 

 submerge any part of it in water.

 

 

temperature being measured is

outside of the range of the instrument,

Batterie okay. 

Messungen 

möglich

Batterie schwach, 

Batterie sollte 

ersetzt werden, 

Messungen 

möglich

Batterie leer, 

 

Messungen nicht 

mehr möglich, 

Ergebnisse könnten 

falsch sein.

Hi oder LO erscheint bei Messerge-

bnissen außerhalb des Messbereiches. 

Hi bei Temperaturen über +500°C und 

Lo bei Temperaturen unter -33°C.

IR9-PEN Infrared Thermometer 

Operating Instructions 

Congratulate on your selection of this IR9-PEN thermometer. The 

IR9-PEN thermometer is a non-contact infrared thermometer with 

circle laser.   

Top view 

Side view 

Operation:   

1.

Press ‘Meas.’ Button, directly to power IR9-PEN on and start 

measurement.   

2.

Release finger to hold the last reading and see the last, 

Maximum and Minimum readings every 2 seconds in turn.   

Hold        Maximum          Minimum 

o

C /

 o

F Measurement Unit Conversion 

Take off battery cover and use a sticker to shift the scale switch     

LCD ERROR MESSAGES

IR9-PEN incorporates visual diagnostic messages as follows: 

‘Hi’  or  ’Lo’  is  displayed  when  the 

‘Hi’  when  higher  than  +500

  o

  o

F

and  ‘Lo’  when lower than 

 -33

 o

C (-27

 o

F).

‘Er2’ is displayed when the IR9-PEN is exposed to 

rapid changes in the ambient temperature.   

‘Er1’ is displayed when the ambient temperature of 

the IR9-PEN exceeds –10

 o

C (14

o

F) OR +50

 o

C

(122

 o

F). In both cases you should allow plenty of 

time (minimum 30 minutes ) for the IR9-PEN to 

stabilize to the working/room temperature 0 to +40

o

C (32 to 104

 o

F)

.

For all other error messages it is necessary to reset 

the IR9-PEN. To reset the IR9-PEN, turn the 

instrument off, remove the battery and wait for a 

minimum of one minute, reinsert the battery and 

turn on. If the error message remains please contact 

the Service Department  for further assistance. 

BATTERIES 

The IR9-PEN incorporates visual low battery indication as follows: 

‘Battery OK’: 

measurements 

  are possible 

‘Battery Low’: battery 

needs to be replaced, 

measurements are 

possible

‘Battery Exhausted’:

measurements are 

not possible 

BATTERY REPLACEMENT 

When the ‘Low Battery’ icon indicates the battery is low, the  

battery should  be  replaced  immediately  with  a  AA  battery,

The battery is located under the battery cover at the rear of

. Slide the battery cover out     

. Take  off  the  old  battery  and  put  a  new  AA  battery  into 

battery compartment. 

  WARNING

. The Temp function is not inte nded for taking measurements 

that require for Live & Die decision or Safety related 

applications.

. The temperature reading of IR9-PEN indicates only the 

surface temperature, and not the actual temperature of an 

object. 

. Keep the unit dry and away fr om any liquids and direct 

sunlight or a very hot object. (Getting too close to a very hot 

object will creates the risk of burn injury, please be careful.) 

. Always make sure the probe lens is clean without any damage. 
. Dispose of used battery promptly and keep away from 

children. 

STORAGE & CLEANING 

SPECIFICATION 

Range 

-33 to +500

o

C(-27 to +932

 o

F)

Resolution 

199.9 to -9.9 

o

C /

 o

F: 0.1 

o

C /

 o

F

 Above 200 or below-10 

o

C /

 o

F F: 1 

o

C /

 o

F

Response Time 

1 second 

Infrared Accuracy  

+

2.0% of reading or +

2.0

 o

C which one is 

greater 

Field of View (D:S)  

   

Emissivity    

0.95 fixed 

Battery Life   

30 hours continuous use 

Operation temp. and 

humidity  

 0 to 50

 o

C; < 90RH%        

Storage temp. and 

humidity  

 10 to 40

 o

C ; < 65 RH% 

Battery     

AA   

Display   

LCD 

Dimensions  

150l(L) x 25(W) x 27(H) mm 

Weight   

74grams including battery 

GUARANTED 

The IR9-PEN is guaranteed for a period of one year from the date 

of  purchase  against  mechanical  and  electrical  manufacturing 

defects. There  are  no  user  serviceable  parts  inside  the   

instrument. Any attempted repair by unauthorized  persons voids 

o

 o

F  

C

(932 )

the IR9-PEN.   

9:1 optics ratio

the warranty.

Always keep the reflect mirror and lens clean. 

 

If it is dirty, use a wet soft cloth or wet cotton

swab to clean it softly and make the lens is 

 fully dry before using the IR9 - PEN .

 

 

IR9 -PEN is not waterproof, do NOT 

 

 submerge any part of it in water.

 

 

temperature being measured is

outside of the range of the instrument,

IR9-PEN Infrared Thermometer 

Operating Instructions 

Congratulate on your selection of this IR9-PEN thermometer. The 

IR9-PEN thermometer is a non-contact infrared thermometer with 

circle laser.   

Top view 

Side view 

Operation:   

1.

Press ‘Meas.’ Button, directly to power IR9-PEN on and start 

measurement.   

2.

Release finger to hold the last reading and see the last, 

Maximum and Minimum readings every 2 seconds in turn.   

Hold        Maximum          Minimum 

o

C /

 o

F Measurement Unit Conversion 

Take off battery cover and use a sticker to shift the scale switch     

LCD ERROR MESSAGES

IR9-PEN incorporates visual diagnostic messages as follows: 

‘Hi’  or  ’Lo’  is  displayed  when  the 

‘Hi’  when  higher  than  +500

  o

  o

F

and  ‘Lo’  when lower than 

 -33

 o

C (-27

 o

F).

‘Er2’ is displayed when the IR9-PEN is exposed to 

rapid changes in the ambient temperature.   

‘Er1’ is displayed when the ambient temperature of 

the IR9-PEN exceeds –10

 o

C (14

o

F) OR +50

 o

C

(122

 o

F). In both cases you should allow plenty of 

time (minimum 30 minutes ) for the IR9-PEN to 

stabilize to the working/room temperature 0 to +40

o

C (32 to 104

 o

F)

.

For all other error messages it is necessary to reset 

the IR9-PEN. To reset the IR9-PEN, turn the 

instrument off, remove the battery and wait for a 

minimum of one minute, reinsert the battery and 

turn on. If the error message remains please contact 

the Service Department  for further assistance. 

BATTERIES 

The IR9-PEN incorporates visual low battery indication as follows: 

‘Battery OK’: 

measurements 

  are possible 

‘Battery Low’: battery 

needs to be replaced, 

measurements are 

possible

‘Battery Exhausted’:

measurements are 

not possible 

BATTERY REPLACEMENT 

When the ‘Low Battery’ icon indicates the battery is low, the  

battery should  be  replaced  immediately  with  a  AA  battery,

The battery is located under the battery cover at the rear of

. Slide the battery cover out     

. Take  off  the  old  battery  and  put  a  new  AA  battery  into 

battery compartment. 

  WARNING

. The Temp function is not inte nded for taking measurements 

that require for Live & Die decision or Safety related 

applications.

. The temperature reading of IR9-PEN indicates only the 

surface temperature, and not the actual temperature of an 

object. 

. Keep the unit dry and away fr om any liquids and direct 

sunlight or a very hot object. (Getting too close to a very hot 

object will creates the risk of burn injury, please be careful.) 

. Always make sure the probe lens is clean without any damage. 
. Dispose of used battery promptly and keep away from 

children. 

STORAGE & CLEANING 

SPECIFICATION 

Range 

-33 to +500

o

C(-27 to +932

 o

F)

Resolution 

199.9 to -9.9 

o

C /

 o

F: 0.1 

o

C /

 o

F

 Above 200 or below-10 

o

C /

 o

F F: 1 

o

C /

 o

F

Response Time 

1 second 

Infrared Accuracy  

+

2.0% of reading or +

2.0

 o

C which one is 

greater 

Field of View (D:S)  

   

Emissivity    

0.95 fixed 

Battery Life   

30 hours continuous use 

Operation temp. and 

humidity  

 0 to 50

 o

C; < 90RH%        

Storage temp. and 

humidity  

 10 to 40

 o

C ; < 65 RH% 

Battery     

AA   

Display   

LCD 

Dimensions  

150l(L) x 25(W) x 27(H) mm 

Weight   

74grams including battery 

GUARANTED 

The IR9-PEN is guaranteed for a period of one year from the date 

of  purchase  against  mechanical  and  electrical  manufacturing 

defects. There  are  no  user  serviceable  parts  inside  the   

instrument. Any attempted repair by unauthorized  persons voids 

o

 o

F  

C

(932 )

the IR9-PEN.   

9:1 optics ratio

the warranty.

Always keep the reflect mirror and lens clean. 

 

If it is dirty, use a wet soft cloth or wet cotton

swab to clean it softly and make the lens is 

 fully dry before using the IR9 - PEN .

 

 

IR9 -PEN is not waterproof, do NOT 

 

 submerge any part of it in water.

 

 

temperature being measured is

outside of the range of the instrument,

IR9-PEN Infrared Thermometer 

Operating Instructions 

Congratulate on your selection of this IR9-PEN thermometer. The 

IR9-PEN thermometer is a non-contact infrared thermometer with 

circle laser.   

Top view 

Side view 

Operation:   

1.

Press ‘Meas.’ Button, directly to power IR9-PEN on and start 

measurement.   

2.

Release finger to hold the last reading and see the last, 

Maximum and Minimum readings every 2 seconds in turn.   

Hold        Maximum          Minimum 

o

C /

 o

F Measurement Unit Conversion 

Take off battery cover and use a sticker to shift the scale switch     

LCD ERROR MESSAGES

IR9-PEN incorporates visual diagnostic messages as follows: 

‘Hi’  or  ’Lo’  is  displayed  when  the 

‘Hi’  when  higher  than  +500

  o

  o

F

and  ‘Lo’  when lower than 

 -33

 o

C (-27

 o

F).

‘Er2’ is displayed when the IR9-PEN is exposed to 

rapid changes in the ambient temperature.   

‘Er1’ is displayed when the ambient temperature of 

the IR9-PEN exceeds –10

 o

C (14

o

F) OR +50

 o

C

(122

 o

F). In both cases you should allow plenty of 

time (minimum 30 minutes ) for the IR9-PEN to 

stabilize to the working/room temperature 0 to +40

o

C (32 to 104

 o

F)

.

For all other error messages it is necessary to reset 

the IR9-PEN. To reset the IR9-PEN, turn the 

instrument off, remove the battery and wait for a 

minimum of one minute, reinsert the battery and 

turn on. If the error message remains please contact 

the Service Department  for further assistance. 

BATTERIES 

The IR9-PEN incorporates visual low battery indication as follows: 

‘Battery OK’: 

measurements 

  are possible 

‘Battery Low’: battery 

needs to be replaced, 

measurements are 

possible

‘Battery Exhausted’:

measurements are 

not possible 

BATTERY REPLACEMENT 

When the ‘Low Battery’ icon indicates the battery is low, the  

battery should  be  replaced  immediately  with  a  AA  battery,

The battery is located under the battery cover at the rear of

. Slide the battery cover out     

. Take  off  the  old  battery  and  put  a  new  AA  battery  into 

battery compartment. 

  WARNING

. The Temp function is not inte nded for taking measurements 

that require for Live & Die decision or Safety related 

applications.

. The temperature reading of IR9-PEN indicates only the 

surface temperature, and not the actual temperature of an 

object. 

. Keep the unit dry and away fr om any liquids and direct 

sunlight or a very hot object. (Getting too close to a very hot 

object will creates the risk of burn injury, please be careful.) 

. Always make sure the probe lens is clean without any damage. 
. Dispose of used battery promptly and keep away from 

children. 

STORAGE & CLEANING 

SPECIFICATION 

Range 

-33 to +500

o

C(-27 to +932

 o

F)

Resolution 

199.9 to -9.9 

o

C /

 o

F: 0.1 

o

C /

 o

F

 Above 200 or below-10 

o

C /

 o

F F: 1 

o

C /

 o

F

Response Time 

1 second 

Infrared Accuracy  

+

2.0% of reading or +

2.0

 o

C which one is 

greater 

Field of View (D:S)  

   

Emissivity    

0.95 fixed 

Battery Life   

30 hours continuous use 

Operation temp. and 

humidity  

 0 to 50

 o

C; < 90RH%        

Storage temp. and 

humidity  

 10 to 40

 o

C ; < 65 RH% 

Battery     

AA   

Display   

LCD 

Dimensions  

150l(L) x 25(W) x 27(H) mm 

Weight   

74grams including battery 

GUARANTED 

The IR9-PEN is guaranteed for a period of one year from the date 

of  purchase  against  mechanical  and  electrical  manufacturing 

defects. There  are  no  user  serviceable  parts  inside  the   

instrument. Any attempted repair by unauthorized  persons voids 

o

 o

F  

C

(932 )

the IR9-PEN.   

9:1 optics ratio

the warranty.

Always keep the reflect mirror and lens clean. 

 

If it is dirty, use a wet soft cloth or wet cotton

swab to clean it softly and make the lens is 

 fully dry before using the IR9 - PEN .

 

 

IR9 -PEN is not waterproof, do NOT 

 

 submerge any part of it in water.

 

 

temperature being measured is

outside of the range of the instrument,

IR9-PEN Infrared Thermometer 

Operating Instructions 

Congratulate on your selection of this IR9-PEN thermometer. The 

IR9-PEN thermometer is a non-contact infrared thermometer with 

circle laser.   

Top view 

Side view 

Operation:   

1.

Press ‘Meas.’ Button, directly to power IR9-PEN on and start 

measurement.   

2.

Release finger to hold the last reading and see the last, 

Maximum and Minimum readings every 2 seconds in turn.   

Hold        Maximum          Minimum 

o

C /

 o

F Measurement Unit Conversion 

Take off battery cover and use a sticker to shift the scale switch     

LCD ERROR MESSAGES

IR9-PEN incorporates visual diagnostic messages as follows: 

‘Hi’  or  ’Lo’  is  displayed  when  the 

‘Hi’  when  higher  than  +500

  o

  o

F

and  ‘Lo’  when lower than 

 -33

 o

C (-27

 o

F).

‘Er2’ is displayed when the IR9-PEN is exposed to 

rapid changes in the ambient temperature.   

‘Er1’ is displayed when the ambient temperature of 

the IR9-PEN exceeds –10

 o

C (14

o

F) OR +50

 o

C

(122

 o

F). In both cases you should allow plenty of 

time (minimum 30 minutes ) for the IR9-PEN to 

stabilize to the working/room temperature 0 to +40

o

C (32 to 104

 o

F)

.

For all other error messages it is necessary to reset 

the IR9-PEN. To reset the IR9-PEN, turn the 

instrument off, remove the battery and wait for a 

minimum of one minute, reinsert the battery and 

turn on. If the error message remains please contact 

the Service Department  for further assistance. 

BATTERIES 

The IR9-PEN incorporates visual low battery indication as follows: 

‘Battery OK’: 

measurements 

  are possible 

‘Battery Low’: battery 

needs to be replaced, 

measurements are 

possible

‘Battery Exhausted’:

measurements are 

not possible 

BATTERY REPLACEMENT 

When the ‘Low Battery’ icon indicates the battery is low, the  

battery should  be  replaced  immediately  with  a  AA  battery,

The battery is located under the battery cover at the rear of

. Slide the battery cover out     

. Take  off  the  old  battery  and  put  a  new  AA  battery  into 

battery compartment. 

  WARNING

. The Temp function is not inte nded for taking measurements 

that require for Live & Die decision or Safety related 

applications.

. The temperature reading of IR9-PEN indicates only the 

surface temperature, and not the actual temperature of an 

object. 

. Keep the unit dry and away fr om any liquids and direct 

sunlight or a very hot object. (Getting too close to a very hot 

object will creates the risk of burn injury, please be careful.) 

. Always make sure the probe lens is clean without any damage. 
. Dispose of used battery promptly and keep away from 

children. 

STORAGE & CLEANING 

SPECIFICATION 

Range 

-33 to +500

o

C(-27 to +932

 o

F)

Resolution 

199.9 to -9.9 

o

C /

 o

F: 0.1 

o

C /

 o

F

 Above 200 or below-10 

o

C /

 o

F F: 1 

o

C /

 o

F

Response Time 

1 second 

Infrared Accuracy  

+

2.0% of reading or +

2.0

 o

C which one is 

greater 

Field of View (D:S)  

   

Emissivity    

0.95 fixed 

Battery Life   

30 hours continuous use 

Operation temp. and 

humidity  

 0 to 50

 o

C; < 90RH%        

Storage temp. and 

humidity  

 10 to 40

 o

C ; < 65 RH% 

Battery     

AA   

Display   

LCD 

Dimensions  

150l(L) x 25(W) x 27(H) mm 

Weight   

74grams including battery 

GUARANTED 

The IR9-PEN is guaranteed for a period of one year from the date 

of  purchase  against  mechanical  and  electrical  manufacturing 

defects. There  are  no  user  serviceable  parts  inside  the   

instrument. Any attempted repair by unauthorized  persons voids 

o

 o

F  

C

(932 )

the IR9-PEN.   

9:1 optics ratio

the warranty.

Always keep the reflect mirror and lens clean. 

 

If it is dirty, use a wet soft cloth or wet cotton

swab to clean it softly and make the lens is 

 fully dry before using the IR9 - PEN .

 

 

IR9 -PEN is not waterproof, do NOT 

 

 submerge any part of it in water.

 

 

temperature being measured is

outside of the range of the instrument,

Er2 erscheint wenn das Messgerät selbst sehr 

heftigen Temperaturänderungen in der Umgebung 

ausgesetzt wird.

Er1 erscheint wenn die Umgebungstemperatur tiefer 

als -10°C oder höher als 50°C ist.

Er3 Bei allen anderen Fehlern, die zu Gerätestörun

-

gen führen wird Er3 angezeigt. Um das Gerät zurück

-

zusetzen (RESET), bitte Batterie für mindestens eine 

Minute entfernen und wieder einsetzen.

Technische Informationen

Technische Daten

Messbereich:

-33 ... +500°C (-27 ... +932°F)

Auflösung:

199.9 ... -9.9°C/°F: 0.1°C/°F

Oberhalb von 200°C oder unter 

-10°C/°F F: 1°C/°F

Ansprechzeit:

1 Sekunde

Optische Auflösung:

9:1

Emissionsgrad:

0.95 nicht einstellbar

Batterielebensdauer:

30 Stunden im Dauereinsatz 

Arbeitsbereich:

Temperatur / Feuchtigkeit

0 ... 50°C / < 90RH%

Lagerbedingungen:

Temperatur / Feuchtigkeit

10 ... 40°C / < 65 RH%

Batterie:

1 x AA-1,5 Volt

Display:

LCD

Abmessungen:

150 (L) x 25 (W) x 27 (H) mm

Gewicht:

74 g inkl. Batterie

Summary of Contents for 5020-0470

Page 1: ...www dostmann electronic de 5020 0470 Bedienungsanleitung 2 8 Instruction for Use 9 15 ScanTemp 470 Infrarot Thermometer Infrared thermometer ...

Page 2: ...cht wenn am Sensor oder am Gehäuse selbstAnzeichen von Beschädigungen erkennbar sind Sollte das Gerät tatsächlich beschädigt sein versuchen Sie bitte nicht es selbst zu reparie ren In diesem Fall setzen Sie sich bitte mit ihrem Händler in Verbindung Batterien nicht ins Feuer werfen kurzschließen auseinander nehmen oder aufladen Explosionsgefahr Batterien enthalten gesundheitsschädliche Säuren Um e...

Page 3: ...y 2 seconds in turn Hold Maximum Minimum o C o F Measurement Unit Conversion Take off battery cover and use a sticker to shift the scale switch Th W ba Th S T b o C o F th Side view Operation 1 Press Meas Button directly to power IR9 PEN on and start measurement 2 Release finger to hold the last reading and see the last Maximum and Minimum readings every 2 seconds in turn Hold Maximum Minimum o C ...

Page 4: ...er the battery cover at the rear of Slide the battery cover out Take off the old battery and put a new AA battery into battery compartment R R R I F E B O h S h B D D W Th of de ins o C o F the IR9 PEN A If s fu IR s Side view Operation 1 Press Meas Button directly to power IR9 PEN on and start measurement 2 Release finger to hold the last reading and see the last Maximum and Minimum readings ever...

Page 5: ...etzlich verpflichtet gebrauchte Batterien und Akkus zur umweltgerechten Entsorgung beim Handel oder entsprechenden Sammelstellen gemäß nationaler oder lokaler Bestimmungen abzugeben Dieses Gerät ist entsprechend der EU Richtlinie über die Entsorgung von Elektro und Elektronik Altgeräten WEEE gekennzeichnet Dieses Produkt darf nicht mit dem Hausmüll entsorgt werden Der Nutzer ist verpflichtet das A...

Page 6: ... piece of soft cloth Do not allow any liquid into the interior of the device Do not use the device when the instrument or a part of the instrument is damaged In case of a defective devise please contact your distributor Batteries must not be thrown into a fire short circuited taken apart or recharged Risk of explosion Batteries contain harmful acids Low batteeries should be chan ged as soon as pos...

Page 7: ...porates visual diagnostic messages as follows Battery changing When the Low Battery icon indicates the battery is low the battery should be replaced immediately with a AA battery The battery is located under the battery cover at the rear of the thermometer Slide the battery cover out Take off the old battery and put a newAAbattery into battery compartment Battery status indicator Side view Operati...

Page 8: ... AA battery into battery compartment Always make sure the probe lens is clean without any damage Dispose of used battery promptly and keep away from children STORAGE CLEANING SPECIFICATION Range 33 to 500 o C 27 to 932 o F Resolution 199 9 to 9 9 o C o F 0 1 o C o F Above 200 or below 10 o C o F F 1 o C o F Response Time 1 second Infrared Accuracy 2 0 of reading or 2 0 o C which one is greater Fie...

Page 9: ... 08 90 Fax 49 0 93 42 3 08 94 E Mail info dostmann electronic de Internet www dostmann electronic de Technische Änderungen vorbehalten Nachdruck auch auszugsweise untersagt Stand01 1602CHB DOSTMANN electronic GmbH Technical changes reserved Reproduction is prohibited in whole or part Stand01 1602CHB DOSTMANN electronic GmbH ...

Reviews: