Doser DM 4 A Operating Instructions Manual Download Page 9

 

 

Customer Specific Linearisation Curves: 

 

It can be programmed up to 10 customer specific linearisation curves. Each curve is programmable with up to 8 
points. The curves are stored in a EEPROM, the curves will not get lost through changing of the battery: For 
changing the linearisation curves you need the PC interface cable and the software DMI-Controller.. The customer 
specific curves can be selected with code 30. The material group is the “spec. curves”. 

 

Basic Calibration: 

 

The moisture meter can be tested andcalibrated with two test modules.   

material group  

test module   rated value  

BD2 (A0)  

PE05  

37±2%  

BD2 (B0)  

PE05  

10±1%  

HD5 (H0)  

PE30  

12±1%  

PD2 (P0)  

PE30  

8,5 ± 0,5%  

 

 

For changing the calibration you need the PC interface cable and the software DMI-Controller.   

Summary of Contents for DM 4 A

Page 1: ...Feuchtigkeitsmessger te Moisture Meter Humidim tre...

Page 2: ...r wood building materials paper and cardboard Description The electronic moisture meter DM4A is used to determine in a matter of seconds the moisture in materials The average moisture down to a depth...

Page 3: ...he meter works in accordance with the principle of an opened plate capacitor The capacity of the capacitor depends on the material dielectric constant of the material in between the plates Compared wi...

Page 4: ......

Page 5: ...ow buttons and confirm it with OK The following parameters are changeable code 11 storage adjustments OFF ON single ON auto ON single single values will be stored by pressing the OK button ON auto All...

Page 6: ......

Page 7: ......

Page 8: ...higher depth The measuring result is an average moisture value With our moisture meter it s possible to observe the drying process If at the same place after several days the moisture result does not...

Page 9: ...nearisation curves you need the PC interface cable and the software DMI Controller The customer specific curves can be selected with code 30 The material group is the spec curves Basic Calibration The...

Page 10: ...or higher moisture content Attention Frames may also causes higher moisture readings Therefore the instrument also can be used to detect frames Moisture measurement in plastic boates Caused through Os...

Page 11: ...Attention Frames or moutings may also causes higher moisture readings Therefore the moisture meter also can be used to detect frames or mountings at the opposite side of the wall...

Page 12: ...checking and calibrating the moisture meter state the type of the meter plastic case for better protection different types of moisture meters special calibration of the moisture meter interface cable...

Page 13: ...roducts and their uses They should not be taken to imply special characteristics or suitability for any specific purpose other than those stated We constantly work to improve our products and reserve...