Customer Specific Linearisation Curves:
It can be programmed up to 10 customer specific linearisation curves. Each curve is programmable with up to 8
points. The curves are stored in a EEPROM, the curves will not get lost through changing of the battery: For
changing the linearisation curves you need the PC interface cable and the software DMI-Controller.. The customer
specific curves can be selected with code 30. The material group is the “spec. curves”.
Basic Calibration:
The moisture meter can be tested andcalibrated with two test modules.
material group
test module rated value
BD2 (A0)
PE05
37±2%
BD2 (B0)
PE05
10±1%
HD5 (H0)
PE30
12±1%
PD2 (P0)
PE30
8,5 ± 0,5%
For changing the calibration you need the PC interface cable and the software DMI-Controller.