Doser DM 4 A Operating Instructions Manual Download Page 8

 

We recommend carrying out regular periodical controlling check measurements, as different local circumstances 
might need different material adjustments. (recommendation: controlling measurements by oven 

 

Screed Moisture Measuring: 

 

For accurate determination of moisture in different kind of screed we reccomend the instrument G812 together 
with the test module PE05. On intrest, please contact:   

DNS-Denzel Natursteinschutz GmbH 

• 

Am Wasserturm 5 

• 

73104 Börtlingen; Germany Tel: +49 (0) 7161 959 

336 

• 

Fax: +49 (0) 7161 959 337 

• 

[email protected] 

• 

www.dns-denzel.de The DM4A can be used for 

measuring moisture differences   
1.

 

1. 

the measurements are non damaging   

2.

 

2. 

the measurements are very quickly done   

3.

 

3. 

if the measured values are too high time sensitive exact measurements e.g. the oven drying 

method, are superfluous.   
 
Even in one room great differences of the moisture in the fooring are possible. Therefore non damaging 
measurements are necessary to find out the critical places for an exact time sensitive measurement as e.g. the 
oven drying method.   

The number of such damaging measurements are then minimised.   

The result of our capacitive moisture meter depends on differences in density, mixtures and surface roughness. It 
depends also from the moisture profile. The influence to the moisture gets smaller for higher depth. The measuring 
result is an average moisture value.   

With our moisture meter it’s possible to observe the drying process. If at the same place after several days the 
moisture result does not change, the moisture is in balance with the air humidity. In this case also at high 
moisture readings an exact comparing measurement e.g. with the oven drying method is recommended. 

 

Oven Drying Method: 

 

The oven drying method is the most acurate way to measure the material moisture in 
wood (ISO 3130-1975), building materials and paper (DIN ISO 287).   

We recommend this for testing and calibrating of all electronic moisture meters.   

Short description:   
1.

 

1. 

For measuring the weights we recommend a balance with an measuring range of 200g and an 

accuracy of 0,01g   
2.

 

2. 

For drying you need an oven with adjustable temperatures of 40, 100 and 104°C   

3.

 

3. 

Take a probe from wood with a sharp saw, avoid edge parts. For building materials take a probe 

with a with a sharp chisel to a depth of approx 3cm. the probe should be at least 20g   
4.

 

4. 

It is very important to take the weight of the first probe immediately, as air humidity may change 

the moisture content. Name of the first weight: wet weight (WW)   
5.

 

5. 

The probe must be dryed in the oven until the weight is constant. The maximum drying 

temperatures: for wood: 104 °C (ISO 3130-1975) for paper 100°C (DIN ISO 287) for e.g. concrete flooring:100 °C 
for anhydride flooring: 40°C   
6.

 

7. 

The name of the dry weight is DW.   

7.

 

8. The moisture content is calculated with the formulas:   

 

Summary of Contents for DM 4 A

Page 1: ...Feuchtigkeitsmessger te Moisture Meter Humidim tre...

Page 2: ...r wood building materials paper and cardboard Description The electronic moisture meter DM4A is used to determine in a matter of seconds the moisture in materials The average moisture down to a depth...

Page 3: ...he meter works in accordance with the principle of an opened plate capacitor The capacity of the capacitor depends on the material dielectric constant of the material in between the plates Compared wi...

Page 4: ......

Page 5: ...ow buttons and confirm it with OK The following parameters are changeable code 11 storage adjustments OFF ON single ON auto ON single single values will be stored by pressing the OK button ON auto All...

Page 6: ......

Page 7: ......

Page 8: ...higher depth The measuring result is an average moisture value With our moisture meter it s possible to observe the drying process If at the same place after several days the moisture result does not...

Page 9: ...nearisation curves you need the PC interface cable and the software DMI Controller The customer specific curves can be selected with code 30 The material group is the spec curves Basic Calibration The...

Page 10: ...or higher moisture content Attention Frames may also causes higher moisture readings Therefore the instrument also can be used to detect frames Moisture measurement in plastic boates Caused through Os...

Page 11: ...Attention Frames or moutings may also causes higher moisture readings Therefore the moisture meter also can be used to detect frames or mountings at the opposite side of the wall...

Page 12: ...checking and calibrating the moisture meter state the type of the meter plastic case for better protection different types of moisture meters special calibration of the moisture meter interface cable...

Page 13: ...roducts and their uses They should not be taken to imply special characteristics or suitability for any specific purpose other than those stated We constantly work to improve our products and reserve...

Reviews: