background image

Dissassembly

1.  Loosen two (2) screws (Det. 5) and remove indicator housing 

(Det. 4) from adapter (Det. 6). Remove dial indicator (Det. 1) 
from indicator housing (Det. 4).

2.  Loosen locknut (Det. 19) with wrench (Det. 20) and remove 

range extension (Det. 16, 17 or 18) from gaging head (Det. 
12).

3.  Loosen two (2) screws (Det. 9) and remove spring (Det. 8).
4.  Loosen two (2) screws (Det. 10) and disengage housing as-

sembly (Det. 6). Remove plunger rod (Det. 7).

5.  Remove centralizer (Det. 11) from gaging head (Det. 12).
6.  Remove retainer (Det. 15) from gaging head (Det. 12). Re-

move gaging pin (Det. 14) and gaging transfer (Det. 13).

Setting

1.  Insert appropriate range extension (Det. 16, 17 or 18) into 

gaging head (Det. 12) and adjust range extension to the ap-
proximate desired diameter.

2.  Insert gage into reference master (master ring gage, bore 

gage setmaster or gage block setup) and adjust range exten-

sion until dial indicator reads zero at first 12 o’clock position. 

Lock range extension in place with locknut (Det. 19).

3.  Rock gage forward and back in reference master to insure 

zero reading. Adjust indicator bezel to zero if necessary.

4.  Repeat Step No. 3

Assembly

1.  Insert gaging pin (Det. 14), gaging transfer (Det. 13) and 

retainer (Det. 15) into gaging head (Det. 12).

2.  Place centralizer (Det. 11) over gaging head (Det. 12).
3.  Insert plunger rod (Det. 7) into gaging head (Det. 12). Place 

housing assembly (Det. 6) over plunger rod (Det. 7) onto 
gaging head (Det. 12) and lock in place with two (2) screws 
(Det. 10).

4.  Insert spring (Det. 8) and lock in place with two (2)screws 

(Det. 9).

5.  Insert dial indicator (Det. 1) into indicator housing (Det. 4) 

and place onto housing assembly (Det. 6). Lock in place with 
two (2) screws (Det. 5) insuring proper contact of indicator 
with plunger rod (Det. 7).

6.  Thread lock nut (Det. 19) onto range extension (Det. 16, 17 

or 18) and assemble extension to gaging head (Det. 12).

Dorsey Dial Bore Gage

Model DBL-998

Range 1.00 - 1.50

Item

PT. No.

Part Name

Item

PT. No.

Part Name

Item

PT. No.

Part Name

1

.0001 Indicator

8

1150

Spring

15

1122

Retainer

2

120

Indicator Spring

9

Screw

16

1114-1 Range Extension

3

110

Indicator Contact

10

Screw

17

1114-2 Range Extension

4

104

Indicator Housing

11

1102

Centralizer

18

1114-3 Range Extension

5

Screw

12

1110

Gaging Head

19

1113

Extension nut

6

1145

Housing Assembly

13

1120

Transfer

20

1160

Wrench

7

1147

Plunger Rod

14

1130

Gaging Pin

tel: 845-454-3111