Error Messages:
Error Message:
What to do or check:
Ldg 0
The scale is attempting to zero on power-up.
This message will remain until the scale is stable.
Air currents or vibration may be the cause. If
problem persists, the pcb or load cell may be
damaged.
NOTE: This message will not appear if parameter
Suo = no.
ov
−
Ld
The scale is in overload. The load on the scale
platform exceeds the scale capacity by more than
105%.
Remove excess weight from scale platform.
If problem persists, recalibrate.
If problem still persists, the motherboard or load cell
may need to be replaced.
gS
−
oL
The scale is in gross overload. The load exceeds
the scale rating and might result in damage to the
scale.
Remove excess weight immediately.
If problem persists, recalibrate.
If problem still persists, the motherboard or load cell
will need to be replaced.
Err 0
Load on the scale exceeds 20% upon scale start
up. Remove excess weight. This error only occurs
when the Start Up Zero
Su0
parameter is set to
20
.
Change this parameter to
FS
to allow for automatic
start up zeroing up to 100% of capacity.
If this does not solve the problem, check the load
cell connection.
Ad oF
Verifying load cell wiring connections. Ensure the
load cell connector on the motherboard is seated
properly.
Er EP
The setup parameters loaded in nonvolatile memory
have become corrupted.
Verify scale parameters and calibrate.