EN
Using the refrigerator
HC302D, HC502D
22
7.5
Alarms
A
There are five special cases when alarms occur:
The alarm can be muted and unmuted:
➤
Press the
button.
I
NOTICE!
•
If you are not able to eliminate the problem triggering the alarm,
store the medicinal products somewhere else as soon as possible.
•
In case of power failure, the inner temperature will raise in 2 hours
(hold over time) from 2 °C to 10 °C at an ambient temperature
of 25 °C.
Alarm event
Number of beeps
Duration of the beeps
Temperature out of range
1 beep / 5 sec
1 sec
Power failure
1 beep / 3 sec
0.5 sec
Inner fan is blocked
n.a.
permanent beeping
Door open for > 20 sec
1 beep / 2 sec
1 sec
Connection problem between
display and controller
Door open for > 60 sec
2 beeps / 2 sec
0,5 sec
NOTE
•
If another alarm is triggered the alarm signal is automatically
unmuted.
•
After all alarm events have passed, a muted alarm signal must be
manually reset to unmuted.