― 24 ―
― 23 ―
Contact us for repair.
START
Meter/Test leads damaged?
①
Connect the red plug of the test lead to measuring
terminal and the black one to the COM terminal.
②
Set the function selector to .
③
Press the SELECT button to select the continuity check ( ).
④
Short-circuit between the red and black test pins.
WXUQVRQ"
Beeper sounds?
Finished
Does not turn on.
Does not sound.
Turns on. Sounds
Looks OK
Damaged
*In the case nothing is displayed, check for the battery.
5-2
「
」
(Max. rated input voltage: 1,000 V dc/ac,
,QLWLDOLPSHGDQFHN
$XWRPDWLFPHDVXUHPHQWIRU
・
V under Low Impedance
7KLV $XWR
・
V function automatically selects measurement
IXQFWLRQ RI '&9 $&9 RU 5HVLVWDQFH EDVHG RQ WKH LQSXW
between the test leads.
With no input, the meter displays “Auto” when it is ready.
When a signal out of -1.0 Vdc to 1.5
Vdc or 3 Vac up to the
rated 1,000 V is present, the meter displays the voltage value
in appropriate DC or AC, whichever larger in peak magnitude.
:LWK QR YROWDJH VLJQDO EXW D UHVLVWDQFH EHORZ 0 LV
present, the meter displays the resistance value. When below
WKHFRQWLQXLW\WKUHVKROGWRLVSUHVHQWWKHPHWHU
further gives a continuity beep tone.
$XWR
・
9PRGHLQSXWLPSHGDQFHLQFUHDVHVDEUXSWO\IURPLQLWLDON
WRDIHZKXQGUHGN·VRQKLJKYROWDJHKDUGVLJQDOV´/R=µGLVSOD\VRQ
the LCD to remind the users of being in such low impedance mode.
Peak initial load current, while probing 1000 V ac for example, can be
XSWRP$9[NGHFUHDVLQJDEUXSWO\WRDSSUR[
P$9[NZLWKLQDIUDFWLRQRIDVHFRQG
'RQRWXVH$XWR
・
V mode on circuits that could be damaged by
such low input impedance. Instead, use function selector or
high input impedance voltage modes to minimize loading for such
circuits.
WARNING