13
INSTRUCTIONS FOR CLEANING AND STERILIZATION OF MINILIGHT, MINIMATE,
MINIBRIGHT SYRINGES IN ACCORDANCE WITH UNI EN ISO 17665 REQUIREMENTS
Warning
• Sterilization must be performed even when using the syringe for the first time.
• Inappropriate sterilization is dangerous for patients and operators.
• Do not perform spray sterilization. Do not place in any disinfectant liquids.
• Do not perform cold chemical or hot air sterilization.
• The appointed staff must be specialized and trained.
• Use the disinfectant in accordance with the specifications set by the
manufacturer indicated on the label.
• Do not use chlorine-based solutions.
• When sterilizing more than one piece in one autoclave do not exceed its
maximum load capacity.
Preparation
Remove dirt from the surface using a disposable paper tissue.
Let air and water flow from the syringe for about 30 seconds in order to clean
the internal channels.
Press the button locateci at the bottom of the sleeve and slide the stainless
steel sleeve off the syringe body.
Unscrew the ferrule and remove the tip.
Manual cleansing
Wipe with a disposable paper tissue and with the aid of drinking water to
remove any impurities and dirt
Automatic cleansing
Not available
Manual disinfection
Perform disinfection only with a disposable tissue and with compatible
disinfectants (in accordance with the instructions included in the product label
and technical data sheet). Recommended disinfectants:
• lncidin liquid
• FD 322 Durr
• Mikrozid AF Liquid
Automatic disinfection
Not available
Manual drying
Dry with disposable paper tissue.
Dry with dry, clean and uncontaminated compressed air both internally and
externally until completely dry.
Do not dry with hot air.
Automatic drying
Not available
Maintenance and checking
No particular maintenance is required.
There is no objective term limiting the usage life of the sleeve: check to see if
there are if any damages or signs of wear and tear, replace the part if necessary.
Packaging
Use appropriately sized sterilization packages made of thermoweldable film.
Sterilization
The sleeve and tip are autoclavable.
Sterilize using EN 13060 ISO 17665-1 compliant class-B steam autoclave.
Sterilize with 3 phases fractionated vacuum 134° C +/- 1 °C with 2.13 bar
pressure, 4 minutes wait.
Never exceed 135° C.
The autoclave must be validated.
Preservation
No particular requirements other than keeping them in their sealed and
sterilized package. Keep them in an appropriate environment and out of direct
sunlight and in a dry place, which should have low bioburden where possible.
Summary of Contents for L2-ECO
Page 32: ...32 Cleaning and Disinfection of the Suction Handpieces...
Page 33: ...Cleaning and Disinfection of the Suction Handpieces 33...
Page 53: ...Instructions for use Electric motor EM 12 L Supply hose VE 10...
Page 67: ...Supply hose VE 10 Connection sheat Connection quick release coupling 15...
Page 97: ......
Page 99: ...Instructions for Use Foot control S NW S N2 S N1...
Page 130: ...32 Manufacturer s declaration...
Page 131: ...33 Manufacturer s declaration...
Page 132: ...34 Manufacturer s declaration...
Page 133: ...35 Manufacturer s declaration...
Page 135: ...1 Minilight Installation and use manual...
Page 146: ...12 14 WIRING DIAGRAM 14 1 VERSION 6F 14 2 VERSION L...
Page 161: ...EN 1801V002 7560100003L02 Installation and Operating Instructions Spittoon valve 3...
Page 162: ......
Page 185: ......
Page 186: ......
Page 187: ......
Page 189: ...EN 1712V002 7117100018L30 Installation and Operating Instructions CAS 1 Combi Separator...
Page 190: ......
Page 207: ...7117100018L30 1712V002 17 Installation 9 Service program EN...
Page 223: ......
Page 225: ...EN 1709V004 9000 606 39 30 Installation and Operating Instructions CS 1 Combi Sepamatic...
Page 226: ......
Page 248: ......
Page 249: ......
Page 250: ......
Page 251: ......
Page 253: ......
Page 266: ...EVA USER MANUAL EN Edition 1 0 April 2020 Pag 14 di 14...