2019-08-30 Galaxy massage - User Manual.doc
Page 22 z 25
WARNING: Use of this equipment adjacent to or stacked with other equipment should be avoided because it could result
in improper operation. If such use is necessary, this equipment and the other equipment should be observed to verify that
they are operating normally.
WARNING: Use of accessories, transducers and cables other than those specified or provided by the manufacturer of this
equipment could result in increased electromagnetic emissions or decreased electromagnetic immunity of this equipment
and result in improper operation.
WARNING: The EMISSIONS characteristics of this equipment make it suitable for use in industrial areas and hospitals
(CISPR 11 class A). If it is used in a residential environment (for which CISPR 11 class B is normally required) this
equipment might not offer adequate protection to radio-frequency communication services. The user might need to
take mitigation measures, such as relocating or re-orienting the equipment.
WARNING: Device may be susceptible to electromagnetic disturbances, but Basic Safety and Essential Performance are
maintained.
Guidance and manufacturer’s declaration –
electromagnetic immunity
The equipment
*
is intended for use in the electromagnetic environment specified below. The customer or the user of
the equipment
*
should assure that it is used in such an environment.
IMMUNITY test
IEC 60601
test level
Compliance level
Electromagnetic environment
–
guidance
Electrostatic
discharge (ESD)
IEC 61000-4-2
± 8 kV (
contact)
± 2/4/8/15 kV (air)
± 8 kV (contact)
± 2/4/8/15 kV (air)
Floors should be wood, concrete or
ceramic tile. If floors are covered with
synthetic material, the relative
humidity should be at least 30 %.
Electrical fast
transient/burst
IEC 61000-4-4
±2 kV for power
supply lines
100 kHz
±2 kV for power
supply lines
100 kHz
Mains power quality should be that of a
typical commercial or hospital environment.
Surge
IEC 61000-4-5
± 1 kV line(s) to
line(s)
± 2 kV line(s) to
earth
± 1 kV line(s) to
line(s)
± 2 kV line(s) to
earth
Mains power quality should be that of a
typical commercial or hospital environment.
Voltage dips, short
interruptions and
voltage variations
on power supply
input lines
IEC 61000-4-11
0 % U
T
; 0,5 cycle at
0°, 45°, 90°, 135°,
180°, 225°, 270° and
315°
0 % U
T
; 1 cycle and
70 % U
T
; 25/30 cycles
(50/60Hz)
1 phase: at 0°
0 % U
T
; 250/300
cycles (50/60Hz)
0 % U
T
; 0,5 cycle at
0°, 45°, 90°, 135°,
180°, 225°, 270° and
315°
0 % U
T
; 1 cycle and
70 % U
T
; 25/30 cycles
(50/60Hz)
1 pha
se: at 0°
0 % U
T
; 250/300
cycles (50/60Hz)
Mains power quality should be that of a
typical commercial or hospital environment. If
the user of the equipment
*
requires continued
operation during power mains interruptions, it
is recommended that the equipment
*
be
powered from an uninterruptible power supply
or a battery.