8.0. TROUBLESHOOTING 

Failure indication 

Possible cause 

Action to be taken 

The adjustment mechanism 
does not work - the LED is off 

off 

If necessary, take the patient from the 
table  with  the  user's  help.  To  restart 
the  controller,remove  the  mains  plug 
from the mains electrical outlet in the 
room  and  the  access  lock  from  the 
socket  and  wait  for  a  minimum  1h. 
Then  connect  the  table  to  the  mains 
socket  and  if  the  LED  is  on,  you  can 
start working with the table. 

Disconnected power cord 

Check  if  the  power  cord  is  plugged 
into  the  power  supply  socket  of  the 
controller and the AC outlet 

Control mechanism 
interrupts operation - LED is 
on 

Exceeded safe working load 

Reduce the load on the table 

 

If fault symptoms persist, stop using the table, unplug the power cord, remove the lock and contact 

the supplier or the manufacturer. 

9.0. Electromagnetic compatibility 

– Guidance and 

manufacturer’s declaration 

Essential Performance - there are no essential performance characteristics according to the risk assessment. 

*

Table GALAXY 

Guidance and manufacturer’s declaration –

 electromagnetic emissions

 

The equipment

*

 is intended for use in the electromagnetic environment specified below. The customer or the user of 

the equipment

*

 should assure that it is used in such an environment. 

Emissions test 

Compliance 

Electromagnetic environment 

 guidance

 

RF emissions 

CISPR 11 

Group 1 

The equipment

*

 uses RF energy only for its internal function. 

Therefore, its RF emissions are very low and are not likely to cause 
any interference in nearby electronic equipment. 

RF emissions 

CISPR 11 

Class A 

The equipment

*

 is suitable for use in all establishments other than 

domestic and those directly connected to the public low-voltage 

power supply network that supplies buildings used for domestic 
purposes. 

Harmonic emissions 

IEC 61000-3-2 

Class A 

Voltage fluctuations/ 

flicker emissions 
IEC 61000-3-3 

Complies 

Summary of Contents for GALAXY JUPITER S3.Fx

Page 1: ...ease call 1 800 494 3395 or visit djoglobal com For international product support please contact your local dealer or your corresponding DJO Global site You may also contact us internationally at the below email address internationalproductsupport DJOglobal com MEDEN INMED Sp z o o 2 WENEDÓW STREET KOSZALIN 75 847 POLAND ...

Page 2: ...ABLE SET 9 3 4 TRANSPORT AND STORAGE 9 4 GENERAL WARNINGS AND CAUTIONS 10 5 PREPARING THE UNIT FOR USE 11 6 OPERATING THE TABLE 13 6 1 LEVELING THE TABLE 13 6 2 HEIGHT ADJUSTMENT RAISING LOWERING THE BED 13 6 3 ADJUSTMENT OF THE FRONT REAR SECTION INCLINATION DECLINATION ANGLE 15 6 4 ADJUSTING POSTURAL DRAINAGE ANGLE 15 6 5 HEIGHT ADJUSTMENT OF THE SIDES IN THE HEADREST PART 16 6 6 MOVING THE TABL...

Page 3: ...is available from your local DJO Service Dealer or the manufacturer on request This warranty will not be fulfilled if the product is used contrary to its purpose or fails to comply with the terms set in the user manual The manufacturer shall not be liable for the consequences of improper i e against the instructions put forth in this user manual use of the accessories for the table CAUTION This si...

Page 4: ... 70 00 85 Hight adjustement minimum 50 52 1 cm 51 53 1 cm H 19 7 20 5 0 4 in 20 1 20 9 0 4 in H maksimum 101 103 1 cm 96 98 1 cm H 39 8 40 5 0 4 in 37 8 38 6 0 4 in H Table width All sections app 64 cm 25 2 in Duty cycyle 18 min 2 min Continuous short term load 10 max 2 minutes on min 18 minutes off Does not apply to a hydraulic actuator table Lifting capacity 200 kg 440 lbs max safe load while ad...

Page 5: ...stment from 0 to 40 degrees Length of the head section 46 cm 18 1 in Head section declination angle adjustment from 0 to 70 degrees Length of the leg section 155 cm 64 in Picture 1 Inclination declination angle adjustment section length of the sections in GALAXY MERCURY CLASSIC tables 2 4 GALAXY MERCURY S2 Fx S2 Fx H Head section inclination angle adjustment from 0 to 40 degrees Length of the head...

Page 6: ... H Head section inclination angle adjustment from 0 to 40 degrees Length of the head section 46 cm 18 1 in Head section declination angle adjustment from 0 to 70 degrees Length of the leg section 155 cm 61 in Picture 5 Angle adjustment of head section in GALAXY MERCURY E S4 Fx S4 Fx H 2 8 GALAXY JUPITER S5 Fx S5 Fx H Head section inclination angle adjustment from 0 to 40 degrees Length of the head...

Page 7: ...ees Picture 7 Angle adjustment of the head postural drainage leg section sizes of sections in GALAXY SATURN P3 Fx P3 Fx H table 2 10 GALAXY SATURN P5 Fx P5 Fx H Head section inclination angle adjustment from 0 to 40 degrees Length of the head section 46 cm 18 1 in Head section declination angle adjustment from 0 to 70 degrees Length of the middle section 54 cm 21 3 in Leg section inclination angle...

Page 8: ...2 Fx H JUPITER Sx Fx H and Saturn Px Fx H table support equipped with hydraulic height adjustment system F0 with two small wheels and two feet F4 made on four feet with foot lever mechanism of four retractable wheels for easy transport upper frame bed mounting surface with height adjustment electric actuator Sx Fx i Px Fx bed mounting surface with height adjustment gas spring Sx Fx H i Px Fx H pat...

Page 9: ...sts LUX STANDARD FLEXI foot support for the therapist remote control for the height adjustment of the bed only MERCURY CLASSIC cap for the face opening option MERCURY CLASSIC Picture 10 Installation of a paper towel roll and recommended unroll direction 3 3 Table set GALAXY massage table model as per order specs Installation and operation manual 1 pc Additional accessories as per order specs 3 4 T...

Page 10: ...t sit on the movable upholstery of the front section INCORRECT CORRECT Picture 11 Correct and Incorrect loading of the table CAUTION Do not allow liquids coffee tea water detergents and disinfectants to contact the actuator electrical wiring and other components of the table In the case of water damage to electrical wiring or actuator conduit the table should be immediately disconnected from the p...

Page 11: ...he table from the socket CAUTION Unplug the table from the mains socket while performing maintenance servicing or similar actions under the table CAUTION When connecting the electrical power supply cable make sure to protect it from accidental clamping of objects between moving parts of the table The electrical power supply cable should be laid in such a manner so as to prevent the patient or the ...

Page 12: ...sconnected from the controller re connect it as follows Insert the power cord into the power supply socket of the controller until the lock is activated Picture 13 Connecting the power cord Before disconnecting the power cord from the controller make sure that the mains plug is removed from the mains power outlet then press the lock with the screwdriver and pull the power cord out of the power soc...

Page 13: ...cs Contact with such agents can cause stiffening and cracking of the cover material as well as a change in gloss of the upholstery surface This is not covered by the warranty Prepare a disposable bed cover such as paper towels disposable sheets etc for each patient before attempting to use the GALAXY table for treatment procedures The patient should wear comfortable underwear and have their body c...

Page 14: ... with hydraulic actuator the height adjuster can be installed in the lower frame and the foot pedal is connected to the actuator In order to raise the bed to the desired height the control pedal should be pressed cyclically in direction B until the desired height is reached Leaving the lever is done by lifting the control lever towards A Leaving the bed is done as long as the lever of the control ...

Page 15: ...e from the mounting holes in the front and middle sections of the table before Using Postural Drainage traction support In GALAXY SATURN model P3 Fx and P5 Fx there is a possibility of adjustment the Postural Drainage It is done by means of an electric actuator see pic 17 This function is activated by using the DPA control panels located at the bottom of the middle section Hold the left button on ...

Page 16: ...sting the height of the sides in the headrest part To change the position of the sides release the lock knob set the desired height and retighten the lock knob 6 6 Moving the table CAUTION Before changing the location of the table make sure that the mains plug of the power cord is removed from the AC outlet and protected against accidental damage during relocation CAUTION It is forbidden to move t...

Page 17: ...ss lock Not applicable to H models GALAXY tables are equipped with a locking mechanism for access to the height adjustment function see pic 20 The adjustment function is locked when key is removed from slot and LED flashes every0 7 seconds Height adjustment is possible when key is in slot and the LED continuously lit In the MERCURY CLASSIC table pic 21 the lock is held by means of a key placed on ...

Page 18: ...ts of the tables after each use to ensure proper hygiene level CAUTION Before starting the cleaning of the table make sure that the electrical power supply cable has been disconnected from the supply in the room to prevent accidental movement of the table CAUTION Estimated useful life of the device is 7 years After 7 years starting from the device production and its equipment the manufacturer is n...

Page 19: ... which is not covered by the warranty Disinfection with the use of the UV rays does not affect the surface of the upholstery 7 2 Maintenance of the base and mechanical parts Clean the metal parts with a soft damp sponge Wipe the parts dry immediately after cleaning Do not use cleaning agents containing alcohol All movable joints must be lubricated every six months or when experiencing squeaks or o...

Page 20: ... quarter for proper action and for sounds other than during normal operation such as strokes squeaks scuffs Any defects in the operation of the actuator should be reported immediately to DJO for technical service and operation of the table stopped until the cause of the fault has been removed 7 5 Periodic check of electrical safety not applicable to H models The electrical safety check should be p...

Page 21: ...manufacturer 9 0 Electromagnetic compatibility Guidance and manufacturer s declaration Essential Performance there are no essential performance characteristics according to the risk assessment Table GALAXY Guidance and manufacturer s declaration electromagnetic emissions The equipment is intended for use in the electromagnetic environment specified below The customer or the user of the equipment s...

Page 22: ...ified below The customer or the user of the equipment should assure that it is used in such an environment IMMUNITY test IEC 60601 test level Compliance level Electromagnetic environment guidance Electrostatic discharge ESD IEC 61000 4 2 8 kV contact 2 4 8 15 kV air 8 kV contact 2 4 8 15 kV air Floors should be wood concrete or ceramic tile If floors are covered with synthetic material the relativ...

Page 23: ...4 6 3 V 0 15 MHz 80 MHz 6 V in ISM bands between 0 15 MHz and 80 MHz 80 AM at 1 kHz 3 V 0 15 MHz 80 MHz 6 V in ISM bands between 0 15 MHz and 80 MHz 80 AM at 1 kHz WARNING Portable RF communications equipment including peripherals such as antenna cables and external antennas should be used no closer than 30 cm 12 inches to any part of the equipment including cables specified by the manufacturer Ot...

Page 24: ... result of defects in material or workmanship and also excludes damage due to lack of maintenance eg bearings guides etc 9 The warranty does not cover consumable materials and components eg gaskets cables etc 10 DJO Global is not responsible for the safety of the user or the patient during the operation of the installed equipment in case of failure to comply with the supplied installation and oper...

Page 25: ...2019 08 30 Galaxy massage User Manual doc Page 25 z 25 Repairs Registry Users Notes ...

Reviews: