17
Guidelines and manufacturer's declaration – electromagnetic immunity
The model ASG 341 is intended for operation in an electromagnetic environment as specified
below. The customer or the user of the model should ensure that it is used in such an
environment.
Immunity
tests
IEC 60601 –
test level
Conformity
level
Electromagnetic environment –
guidelines
Portable and mobile radio devices should
not be used in closer proximity to the
[device or system] including the cables as
the recommended protective distance
calculated in accordance with the formula
for the respective transmission frequency.
Recommended protective distance:
Conducted HF
disturbances
according to
IEC 61000-4-6
3 Vrms
150 kHz to 80
Mhz
3 Vrms
d = 1.2
Radiated HF
disturbances
according to
IEC 61000-4-3
3 V/m
80 MHz to 2.5
Ghz
3 V/m
d = 1.2 80 MHz to 800 MHz
d = 2.3 800 MHz to 2.5 GHz
Where P is the rated power of the
transmitter in watt (W) according to the
specification of the transmitter
manufacturer and d is the recommended
protective distance in metres (m).
According to an on-site
a
investigation, the
field strength of stationary radio
transmitters is in all frequencies lower than
the conformity level.
b
Disturbances are possible in the vicinity of
devices carrying the following symbol.
NOTE 1 For 80 MHz and 800 MHz, the higher value is applicable.
NOTE 2 These guidelines might not apply to all situations. The spreading of electromagnetic
waves is influenced by absorptions and reflections of buildings, objects and people.
ELECTRIC IMMUNITY
12.0 Information regarding electromagnetic immunity
GB