 WARNING:

 Never attempt troubleshooting when the resident is on or near the bed. Always 

ensure the work area is free from residents and environmental hazards before conducting any 
troubleshooting steps. 

Effect

Possible Cause

Verification

Corrective Action

The bed 

doesn’t work 

at all.

Bed is not plugged 

into wall outlet or 

wall outlet doesn’t 

have power.

Check wall outlet for power by 

plugging in another product, 

such as a lamp, to make sure 

it works.

If outlet isn’t working, repair wall 

outlet. If outlet is working, plug 

bed into outlet. It should fit firmly 

into the outlet.

Damaged power cord. Inspect power cord for cuts, 

fraying or other damage.

Power cord may need 

replacement. Contact Direct 

Supply Equipment & Furnishings. 

Wire connections 

may be loose or 

damaged.

Visually check wire 

connections at the control 

box and at the Y-cable. They 

may be loose.

Reconnect any loose wires and/or 

power cords.

Pendant not  

plugged in.

Check the pendant connection 

first at the side of the bed 

then at the control box for 

loose wires.

Plug pendant in securely, making 

sure the connection holder is 

closed to hold the connector in 

place.

Pendant is not 

working.

Plug a pendant you know 

is working into the bed and 

check functions.

If bed works, replace pendant. 

If bed doesn’t work, continue 

troubleshooting. 

Blown fuse in  

control box.

The control box features time 

delay self-resetting fuses that 

do not need replacement. To 

reset the fuse, unplug the bed 

for 5 minutes and then plug 

back in. The bed should be 

operational.

If unplugging the bed and waiting 

5 minutes does not correct the 

problem, contact Direct Supply 

Equipment & Furnishings.

One or more 

functions 

won’t work.

Wire connections 

may be loose.

Visually check all 

connections at the 

control box and pendant 

connections. They may be 

loose.

Reconnect any loose wires and/or 

power cords.

Pendant is  

not working.

Plug a pendant you know 

is working into the bed and 

check functions.

If bed works, replace pendant. 

If bed doesn’t work, continue 

troubleshooting.

Damaged  

actuator cord.

Inspect actuator cord 

for cuts, fraying or other 

damage.

Actuator cord may need to be 

replaced. Contact Direct Supply 

Equipment & Furnishings.

Faulty actuator.

Disconnect power cord from 

bed that is not functioning. 

Swap a working actuator 

plug with the nonworking 

actuator plug on the control 

box. Reconnect the power and 

test functions on the bed by 

pressing the pendant buttons. 

A faulty actuator will not work 

with any connection port.

If the actuator that was previously 

not working now works, but 

with a different pendant button, 

replace the control box after 

verifying the pendant works. 

If the actuator that previously 

did not work still doesn’t work, 

replace actuator. 

Troubleshooting

Summary of Contents for Panacea 6300

Page 1: ...d from Direct Supply Equipment Furnishings Please read this entire guide carefully and keep it for future reference This guide will provide you with instructions warnings warranty information and othe...

Page 2: ...eadboard and Footboard Installation and Adjustment 16 Assist Device Assembly 17 3 Position Assist Device 18 Width Extension 19 Wall Guard 20 Bed Functions 21 Bed Mobilization and Stabilization 22 Batt...

Page 3: ...s and Symbols NOTE Indicates a helpful tip CAUTION Indicates correct operating or maintenance procedures in order to prevent damage to or destruction of the equipment or other property WARNING Calls a...

Page 4: ...ensure tubing does not become entangled kinked or severed during normal operation of the bed Warning This device includes assist devices Some states may prohibit assist devices or consider assist dev...

Page 5: ...re cord does not become entangled or damaged during normal operation of the bed Keep all moving parts free from obstruction e g blankets sheets heating blankets pads tubing wiring etc which may get ta...

Page 6: ...etainers inward Perform this procedure on both sides of the bed The bed is now at 37 wide and can be used to evacuate resident out of the facility WARNING The device should ONLY be used to transport r...

Page 7: ...to clean if necessary Inspect the resident s bed controls looking for signs of damage where liquids could leak in Test the bed and its controls including resident lockout features to ensure the bed c...

Page 8: ...ISO 7000 1641 Operating instructions Change Auto Contour position Recycle the item in accordance with local regulations Change hi lo position of mattress support platform Intertek s ETL listed mark f...

Page 9: ...nto the control box of bed i e pendants and actuators while mains cable power cord is plugged into the wall outlet Any cords or tubing used on or with this bed MUST be routed and secured properly to e...

Page 10: ...s and options not all accessories and options can be used at the same time Assumes Max Patient Resident Weight 582 lbs 264kg Assist Rail pair 35 lbs 16kg Assist Rail 4 70 lbs 32kg Headboard Footboard...

Page 11: ...ame and assist rails are secured to frame to decrease risk of entrapment On March 10 2006 the U S Food and Drug Administration FDA released guidelines for reducing the risk of hospital bed entrapment...

Page 12: ...System 5 Locking Casters Fall Mat Left and Right Side Pendant Connections Trapeze Free Standing or Bed Mounted Slat Deck Staff Control Panel Heavy Duty Casters Mattress Retainers Beige Powdercoated F...

Page 13: ...endelenburg Reverse Trendelenburg capable No for 3 function Yes for 5 function Mattress length 80 mattress deck length 80 203 cm Mattress length 84 mattress deck length 84 213 cm Mattress length 88 ma...

Page 14: ...rd in it 5 Remove ties from pendant Before plugging the pendant into the bed frame cut the black O ring at the end of the pendant so it will snap into the pendant housing and fit securely into the bed...

Page 15: ...ainers four narrow retainers for the sides of the bed and two wider retainers for the end of the bed Insert the two end mattress retainers in the holes appropriate for the mattress length being used L...

Page 16: ...embly at both ends of the bed frame Place the mounting brackets of the headboard footboard into the slots of the support assembly Slide headboard footboard down until it stops Screw the knobs back in...

Page 17: ...ment with rails may cause injury or death To help prevent entrapment you should follow the FDA s Entrapment Guidance and your healthcare provider s recommendations when choosing a bed mattress and acc...

Page 18: ...th assist devices may cause injury or death To help prevent entrapment you should follow the FDA s Entrapment Guidance and your healthcare provider s recommendation to choose a bed mattress and access...

Page 19: ...The width extension expands by using the push button mechanism Push the expansion button mechanism in and pull out width extension until you reach the desired width of the deck 39 Width Setting 42 Wi...

Page 20: ...on the crossbar under the head of the bed The Wall Guard should now be in place Place the Wall Guard into the holes of the mount underneath the head section of the bed frame Place pins on each side of...

Page 21: ...cal personnel 5 Function Staff Control Pad Operation available only on 5 function bed This controller is located on the footboard 1st Set of Buttons Head deck up and down 2nd Set of Buttons Foot deck...

Page 22: ...leave a bed unattended while the caster locks are not engaged Caster Lock To lock the bed frame push down on all four caster locks with your foot CAUTION Moving the bed while the floor lock or caster...

Page 23: ...bed Battery is not to be used in a damp environment such as outdoors WARNING The battery backup should be used for EMERGENCY PURPOSES ONLY The only compatible battery backup is a DewertAG7 battery DO...

Page 24: ...t expose the battery to heat or open flames Avoid storing it in direct sunlight If the battery leaks and you come into contact with the fluid wash the affected area thoroughly in water and consult a...

Page 25: ...to make sure actuators do not bind at any point throughout their full range of motion Authorized Accessories Inspect all fasteners for looseness and wear Replace or tighten as necessary Ensure welds d...

Page 26: ...nplug power supply cords Ensure all electrical parts motors control boxes wires and pendant are not broken Ensure NO liquids enter electrical components Remove all gross solid contaminants then wash a...

Page 27: ...bed Replace tie wraps to hold power cord to frame Plug mains cable into wall outlet Test function WARNING When routing cables in the medical bed precautions shall be taken to avoid squeezing between...

Page 28: ...mide Solvent Red Pendant Housing PA 765A Rail Release Handle ABS Headboard Footboard Wood Trapeze PVC Steel Nylon Webbing General information The recommended environment for operation and storage tran...

Page 29: ...d wastes throughout the EU To prevent possible harm to the environment or human health from uncontrolled waste disposal recycle it responsibly to promote the sustainable reuse of material resources To...

Page 30: ...bed doesn t work continue troubleshooting Blown fuse in control box The control box features time delay self resetting fuses that do not need replacement To reset the fuse unplug the bed for 5 minute...

Page 31: ...hat the bed may get caught on Reposition the bed or move object causing the obstruction Lower the bed all the way down until all movement has stopped The bed should now be level and remain level when...

Page 32: ...es Harmonic Emissions IEC 61000 3 2 Not Applicable Voltage fluctuations flicker emissions IEC 6100 3 3 Recommended separation distances between portable and mobile RF communications equipment and the...

Page 33: ...fixed RF transmitters a as determined by an electromagnetic site survey should be less than the compliance level in each frequency range b Interference may occur in the vicinity of equipment marked w...

Page 34: ...od identified below whichever is earlier Warranty Period Parts Only Anticipated Usable Device Life Frame 10 years on frame 20 years on welds 5 years Electrical Parts Components 1 year 1 year Anticipat...

Page 35: ...uality Panacea 6300 Bariatric Bed and enjoy outrageous customer service If you have any questions about the bed you have purchased or would like to request warranty service please contact Direct Suppl...

Page 36: ...rect Supply Inc Distributed by Direct Supply Manufacturing Inc 6699 N Industrial Road Milwaukee WI 53223 2010 2015 Direct Supply Manufacturing Inc All Rights Reserved Direct Supply Panacea and all ass...

Reviews: