PRODUCT MAINTENANCE
Maintenance of the instruments and handpieces should be carried out according to the instructions of their manufacturers.
In case of the spitton block version with the connection to thecentral distribution, check the cleanliness of the advance fileter
/strainer/ andtheoperation of the water hardness treatment device (according to the instructions of the manufacturer).
Inspection during the guarantee period
During the warranty period, the user is recommended regularly invite an authorized service technician for preventive
periodical inspection in
3-month intervals.
The inspection is focused on the following
•
check of input filters (with regard to the cleanliness of input media)
•
check of the suction system
•
check of the waste hose
•
providing additional information and practice advice concerning the dental unit
•
check of the proper use and maintenance of the dental unit and its instruments (according to the Instructions for
Use and training)
•
check and/or adjustment of all media (input, setting of turbine pressures and the like)
•
the manufacturer determines therange of above-mentioned work to be
1
to
1,5 hours
•
authorized service technician is obliged to confirm the periodical inspection carried out in the guarantee card.
Inspection and audit /revision/ after the guarantee period:
Periodical inspection of the product must be carried out by an authorized service technician in
6-month intervals,while
the following is recommended to be done:
•
general inspection of the dental unit and functional parts thereof
•
check and adjustment of the working pressures of water and air
•
check of water and air filters in the spittoon block
•
check of the integrity of the electropart and electric wiring (electrical safety)
•
Audit /revision/ of electrical safety
It is carried out according to the regulation of the country in which the dental unit is installed.