![Dionex UltiMate 3000 LCi Quick Installation Manual Download Page 32](http://html.mh-extra.com/html/dionex/ultimate-3000-lci/ultimate-3000-lci_quick-installation-manual_2506103032.webp)
UltiMate
®
3000 Viper™ Capillary Kits for Intelligent LC (LCi) System Solutions
Quick Installation Guides
Page 28
Rev. 02
January 2010
©
2010 Dionex
Table 10:
Detailed description of capillary connections of the Parallel LC setup (dual-stack)
Description of Connections
Capillary Dimensions (ID x L)
PumpLeft outlet – TCC 6-port valve left, port 6
0.18 mm/0.007” x 550 mm
PumpRight outlet – TCC 6-port valve left, port 4
0.18 mm/0.007” x 350 mm
TCC 6-port valve left, port 5 – WPS sampler valve, port 5
RS: 0.13 mm/0.005” x 250 mm
SD: 0.18 mm/0.007” x 250 mm
WPS sampler valve, port 4 – TCC 6-port valve left,
port 2
RS: 0.13 mm/0.005” x 250 mm
SD: 0.18 mm/0.007” x 250 mm
TCC 6-port valve left, port 3 – Column 1 inlet
RS: 0.13 mm/0.005” x 250 mm
SD: 0.18 mm/0.007” x 250 mm
TCC 6-port valve left, port 1 – Column 2 inlet
RS: 0.13 mm/0.005” x 250 mm
SD: 0.18 mm/0.007” x 250 mm
Column 1 outlet – Detector 1 inlet
RS: 0.13 mm/0.005” x 350 mm
SD: 0.18 mm/0.007” x 350 mm
Column 2 outlet – Detector 2 inlet
RS: 0.13 mm/0.005” x 350 mm
SD: 0.18 mm/0.007” x 350 mm
Note:
In case of a single-stack setup of the Parallel LC system, the TCC 6-port valve
left
turns into a TCC 6-port valve
right
. The capillary lengths have to be
adapted as outlined in Table 11.
Table 11:
Detailed description of capillary connections of the Parallel LC setup (single-stack)
Description of Connections
Capillary Dimensions (ID x L)
PumpLeft outlet – TCC 6-port valve left, port 6
0.18 mm/0.007” x 950 mm
PumpRight outlet – TCC 6-port valve left, port 4
0.18 mm/0.007” x 950 mm
TCC 6-port valve left, port 5 – WPS sampler valve, port 5
RS: 0.13 mm/0.005” x 550 mm
SD: 0.18 mm/0.007” x 550 mm
WPS sampler valve, port 4 – TCC 6-port valve left,
port 2
RS: 0.13 mm/0.005” x 550 mm
SD: 0.18 mm/0.007” x 550 mm
TCC 6-port valve left, port 3 – Column 1 inlet
RS: 0.13 mm/0.005” x 250 mm
SD: 0.18 mm/0.007” x 250 mm
TCC 6-port valve left, port 1 – Column 2 inlet
RS: 0.13 mm/0.005” x 250 mm
SD: 0.18 mm/0.007” x 250 mm
Column 1 outlet – Detector 1 inlet
RS: 0.13 mm/0.005” x 350 mm
SD: 0.18 mm/0.007” x 350 mm
Column 2 outlet – Detector 2 inlet
RS: 0.13 mm/0.005” x 350 mm
SD: 0.18 mm/0.007” x 350 mm
Summary of Contents for UltiMate 3000 LCi