2. Locate and eliminate any liquid leaks.
•
Contaminated or incorrect eluent.
Remake the eluent using reagent grade chemicals and ASTM
filtered Type I (18 megohms or 1
µ
S) grade deionized water.
•
Contaminated or degraded sample.
Take appropriate precautions when preparing and storing
samples to prevent contamination and degradation.
•
Column is contaminated.
1. Clean the column (see Column Rejuvenation Procedures,
Technical Note 2R, Document No. 032036).
2. If cleaning is unsuccessful, replace the column.
4
•
Troubleshooting
Doc. 034859-02 9/93
4-7