3 • Operation and Maintenance
Doc. 065125-01 10/06
27
Sampler Options Screen: Syringe Pump Version
When the sample loading pump is the syringe pump, this screen is
displayed:
Select the
Needle Height
.
Select the
Rinse Volume
.
Select the syringe speed for the remaining parameters in the
Syringe
Pump
group box. There are two syringe speeds (fast and slow), and
each speed has 10 levels. The actual speed depends on multiple
factors; for details, refer to the following sections.
“Fast” Syringe Speed
The syringe operates at the “fast” speed for
Aspirate Speed from
Flush Bottle
and
Dispense Speed to Waste Bottle
.
In the default AS-HV configuration, liquid moves through
large
bore
tubing during these operations. When selecting a speed setting, be
sure to take into account the tubing bore.
Summary of Contents for AS-HV
Page 1: ...AS HV Autosampler Operator s Manual Document No 065125 Revision 01 October 2006...
Page 8: ...AS HV Operator s Manual vi Doc 065125 01 10 06...
Page 28: ...AS HV Operator s Manual 20 Doc 065125 01 10 06...
Page 54: ...AS HV Operator s Manual 46 Doc 065125 01 10 06...
Page 66: ...AS HV Operator s Manual 58 Doc 065125 01 10 06...
Page 92: ...AS HV Operator s Manual 84 Doc 065125 01 10 06...
Page 116: ...AS HV Operator s Manual 108 Doc 065125 01 10 06...