9
6.7 SAMPLE WEIGHT PERCENTILE
In this functioning mode, the display shows the weight expressed as a percentile, comparing it with a reference weight
for which the 100 percentile has been already linked.
MATCHING PROCEDURE
-
Put the reference weight on the scale
.
-
Press the
%
key when the “stable” indicator is on.
-
The scale has executed the percentile matching and displays the percentile of the weight on the scale.
-
Unload the scale: the following weighs will be expressed as a percentage in reference to the reference weight.
-
To cancel or execute a new matching operation
, press the
%
key and repeat the operations described above.
6.8 UNIT OF MEASURE CONVERSION
By pressing various times the
MODE
key the conversion into the secondary units of measure takes place; the relative
symbol will turn on in the display (see section 5).
NOTE:
The function is not available if the counting function or the percentile function is enabled.
6.8.1 SECONDARY UNITS OF MEASURE
No
Unit mark
Unit name
Conversion (g)
1
g
gram
2
g
gram (weight x 10, high resolution)
3
ct
net carat
=0.2g
4
lb
lb
=453.59237g
5
oz
ounce
=28.349523125g
6
d
dram
=1.7718451g
7
ozt
troy ounce
=31.1034768g
8
dwt
penny weight
=1.55517384g
9
mom
momme
=3.749996g
10
tl.h
tael twn
=37.49995g
11
tl.c
tael chn
=37.799375g
12
Tl.t
troy tael
=37.4290018g
13
t
tola
=11.6638039g
14
Bt
Bangladeshi tola
1bt=11.6638039g
15
n
newton
NOTE:
It is possible to enable or disable the unit of measure in the “F1 Unt” step, see section 7.2.
7. SET-UP ENVIRONMENT
The scale has various parameters that can be configured in the SET-UP environment.
To enter in the SET-UP environment press during the start-up the
MODE
key during the countdown.
- The display will show the first function, which is "F1 Unt":
- Pressing the
MODE
will cycle through the other functions.
- Pressing
ZERO
will allow you to set the function.
- Use
MODE
to select the value, or set a value using the
%
key and the
PCS
key to move the active digit and then using
the
TARE
key to increment a digit
- Press the
ZERO
key to confirm the modification and pass to the following step or use the
PRINT/M+
key to leave a
parameter unchanged or to escape set-up environment.