Dini Argeo HLD Series Operating Manual Download Page 13

 

14

P CONT   

In  this  transmission  mode  the  instrument  transmits  continuously  the  following  data  format  depending  on  the  functioning 
mode: 
 
Totalisation Mode 

 
 
 

 
Counting Mode 

 
 
 

 
Percentage Mode 

 
 
 

 

 
9. ERROR MESSAGES 

 

ERROR

 

MESSAGES 

DESCRIPTION 

RESOLUTION 

- - - - - - 

Over range 

Remove weight from the scale. 

Err 4 

Zero Setting Error 

Upon  start-up  or  when  the 

ZERO

  key  is  pressed,  the  weight  on  the 

scale exceeds the percentage programmed on the max. capacity. 
Remove the excess weight and retry. 

Err 5 

Keyboard Error.          The keyboard could be damaged. 

Err 6 

Input signal greater 
than 3mV/V 

Indicates  that  the  A/D  converter  value  is  greater  than  the  maximum 
value:   
- Remove the weight from the scale if it’s exceeding. 
- The load cell or the electronics could be damaged. 

 
RECYCLING INSTRUCTION 

The crossed-out wheeled bin on the product means that at the product end of life, it must be taken to 
separate  collection  or  to  the  reseller  when  a  new  equivalent  type  of  equipment  is  purchased.  The 
adequate  differentiated  refuse  collection  in  having  the  product  recycled,  helps  to  avoid  possible 
negative effects on the environment and health and supports the recycling of the materials of which 
the equipment is made. The unlawful disposal of the product by the user will entail fines foreseen by 
the current regulations. 

ST,GS,XXX.XXXXkg

<CR><LF>

 

<CR><LF> 

    

 

 

 

 

GS for gross weight, NT for net weight, XXX.XXXX is the weight 

PCS:XXXXXXXXpcs

<CR><LF>

 

<CR><LF> 

    

 

 

 

 

XXX.XXXX is the Number of pieces 

PERC:XXXXX.XX%

<CR><LF>

 

<CR><LF> 

    

 

 

 

 

XXXXX.XX is the Percentage of weight 

Summary of Contents for HLD Series

Page 1: ...RWAAGE BALANCE POUR LABORATOIRES BILANCIA DA LABORATORIO HLD OPERATING MANUAL Industrial Weighing Systems 9 Richmond St Picton Ontario Canada K0K 2T0 www iwsystems ca e mail info iwsystems ca HLD_V1 0...

Page 2: ...ATION OF THE WEIGHT 8 6 5 1 MANUAL OR AUTOMATIC TOTALIZATION 8 6 6 COUNTING PIECES 8 6 7 SAMPLE WEIGHT PERCENTILE 9 6 8 UNIT OF MEASURE CONVERSION 9 6 8 1 SECONDARY UNITS OF MEASURE 9 7 SET UP ENVIRON...

Page 3: ...y are not in use 2 MAIN TECHNICAL SPECIFICATIONS Model Plate dimensions mm Capacity g Standard Division g HR Division g Linearity g Internal sensitivity g Min APW recommended g HLD150 80mm 150 0 02 0...

Page 4: ...is advisable to completely recharge the battery 12 hours with the first installation of the instrument TO POWER the instrument through the 240 Vac mains put the AC DC power plug in the appropriate so...

Page 5: ...matic tare A secondary function of incrementing the active digit or selecting a value or parameter when in setting mode If pressed for an instant it allows linking a weight to the 100 percentile see s...

Page 6: ...orm a value for its weight will be displayed Press the TARE key to tare the scale when the stable is light The weight that was displayed is stored as the tare value and that value is subtracted from t...

Page 7: ...N The operation of accumulation normally done manually with key PRINT M can be done automatically when the weight is stable In regards to the choice of the totalisation execution mode see step AU sect...

Page 8: ...t name Conversion g 1 g gram 2 g gram weight x 10 high resolution 3 ct net carat 0 2g 4 lb lb 453 59237g 5 oz ounce 28 349523125g 6 d dram 1 7718451g 7 ozt troy ounce 31 1034768g 8 dwt penny weight 1...

Page 9: ...ion 6 8 enabling on or disabling oFF the suggested secondary units of measure 1 The display proposes the first secondary unit of measure to be enabled or disabled 2 Press MODE to change the options an...

Page 10: ...password To enter the menu press the PCS M and TARE keys one after the other The display shows the first parameter P1 Lin IN CASE THE ZONE OF USE IS DIFFERENT FROM THE CALIBRATION ZONE ONE SHOULD 1 Se...

Page 11: ...POINTS By pressing ZERO the instrument shows the A D converter points relative to the weight on the scale Press PRINT M to exit the step P4 AZ ZERO TRACKING By pressing ZERO one sets the divisions num...

Page 12: ...ccumulating the weight the data format is the following When the totals are recalled from the storage the data format is the following Counting Mode In the counting operations when the PRINT M key is...

Page 13: ...t from the scale if it s exceeding The load cell or the electronics could be damaged RECYCLING INSTRUCTION The crossed out wheeled bin on the product means that at the product end of life it must be t...

Reviews: