DOSAGE
46
Once the first formula is finished, the instrument:
- executes the automatic tare
- executes the following formula (in Production Program)
These operations are repeated for all the formula.
At the end, the instrument waits to receive the start of the new dosage, or, in case of various repetitions of
the dosage cycle, restarts automatically with the following cycle.
To pause a dosage under way, one should:
- press the external PAUSE button, which must be connected to the IN2 input of the instrument, or the
button.
- to restart the interrupted dosage, give a START command.
For further details, see section “Pause – momentary interruption of the dosage cycle”
To reset a dosage under way, one should:
- press at length the external PAUSE button or press 2 times the
button, until the display shows
the "RESET DOSAGE" message.
For further details, see section “Dosage cycle reset”.
To set the number of times in which the instrument must repeat the selected formula, one should:
- press the
button and enter the number of repetitions to be executed.
For further details, see section “Setting the number of consecutive ripetition of the active formula”.
6.13.1
Dosage cycle starts
To start the dosage cycle, press for an instant the external INPUT 1 or the START button. In order to start
with the dosage, the weight on the scale:
1)
Must be stable or become stable within a waiting time equal to the value set in the
TIME FOR
MATERIAL FALLING DOWN
parameter (
).
2)
Must be less than the value set in the
ZERO TOLERANCE WEIGHT
parameter (
)
.
3)
Must be greater than
MINIMUM TARE
and less than
MAXIMUM TARE
(if programmed)
.
4)
Must be stable within the time set in the
WAIT STABILITY TIME
parameter (
)
If any one of these condition are not respected an error message is shown on the system status
By pressing for an instant the external INPUT 1 or the START button, the cycle starts again.
6.13.2
Pause --- momentary interruption of the dosage cycle
To pause the dosage, it is necessary to close for an instant the PAUSE input or push the
button.
By carrying out this operation all the instrument’s outputs are disabled and the PAUSE message appears on
the display.
To exit this PAUSE condition, press the
button.