DOSAGE
56
The present weight is lower
than the selected formula
preset minimum tare value
on dosage start
To
check
the
selected
formula minimum tare in the
database management and
change the minimum tare
value or load the scale until
the weight is over the
formula minimum tare
The remote DGTQF can not
be communicated correctly
To check the connection and
the configuration of the
DGTQF and EGTBATCH
The PC com communication
mode is not on 485 when
remote dosage activity is
trying to start
To
set
the
PC
com
communication
mode
working on 485 in the
technical setup
The first activity is not correct
in First free component
formula’s target setting mode
To set the first activity in the
formula as manual dosage
type or manual weight type
The activities number in the
formula is not set correctly
To reset the activities number
and inserted the correct
activity content