14
(3) After the corresponding parameters are selected, press the [Enter] key again
to enter the corresponding parameter modification operation. At this time,
the lower row does not blink, but the parameter content in the upper row blinks,
which can be modified by the four keys [←] [→].
(4) After the modification, press [Enter] to confirm the completion of the setting,
and press [Cancel] to abandon the current item setting and exit to the previous
menu.
Note: For some parameters, users can completely enter their own set values,
while for others, they can only select the built-in fixed paramet
ers through [↑]
[↓], and users can judge whether the parameter values displayed in the upper row
are flashing or not.
5.5.2 F parameter table
Lower row
display
Upper row
display
meaning
[
,,,F00
] [
PP -----
]
Prompt for login password, see section 5.4 for details.
[
,,,F01
]
[
d 0 1 0
]
Grading value (01,02,05,10,20,50,100 optional)
[
,,,F02
]
[
p 0
]
If the number of decimal places (0-4 optional) of the
weight exceeds, the error code [Error ,08] will be
displayed.
[
,,,F03
]
[
3 0 0 9 0
]
The full-scale value of the scale (factory default
value: [,,030090])
[
,,,F04
]
[
r- o x . y
]
0-Cannot be set to zero 1-1% 2-2% 3-5% 4-10%
5
-
20%
6
-
50%
≥
7
-
100%
X
–
auto-zeroing range at power-on (percentage of the full-scale
value of the scale)
Y
–
key zero setting range (percentage of the full scale value of
the scale)
For example, setting it to "2.5" means that the range
of automatic zero setting at power-on is 2%, and the
range of key zero setting is 20%. The factory default
setting is "1.1"
[
,,,F05
]
[
r- a 0 . 5
]
Zero tracking range (setting range: 0.0 ~ 9.9 division
values)
[
,,,F06
]
[
m o d e 0 2
]
Communication mode, see Section 7.1 for details
[
,,,F07
]
[
A d r 0 1
]
Communication address (1-26 optional), in continuous
transmission mode, it means to select the content to
be sent
Quick reference table of group F parameters (continued 1)