5.4 Operation steps of password login
The specific operation steps of password login are as follows:
button
Lower row
display
Upper row
display
meaning
【
Fxx
】
[
,,
F00
] [
pp -----
]
In the weighing display state, press and hold
the [Fxx] key for a long time, and the lower
row will flash. If you have logged in, the
lower row displays [F01]. If you need to
switch the login level, you can press the
【↓ 】
key and select the parameter F00. If
you enter the current password again, you
will enter the password change function. See
Section 7.5 for details.
[input]
[
,,
F00
] [
pp -----
] The top row flashes the rightmost digit.
directio
n key
[
,,
F00
] [
pp ----1
]
Enter the corresponding password, factory
default value: user password [-1],
administrator
password
[59565],
manufacturer password, dynamic uncertainty.
[input]
[
,,
F00
] [
,,User-1
]
According to the unused password, the
corresponding login prompt is displayed,
which indicates that the login is
successful. If the input password is wrong,
[,err,,] will be displayed, and after 1
second, it will automatically exit and enter
the weighing display state.
Table 5.4 Specific Operation Steps of Password Login
Note: After the power is turned on again, you will automatically log out of
the login state. If you need to operate the corresponding items again, you need
to log in again. If you need to log out of the logged-in state, you can also take
the initiative to power off and restart the instrument.
Note: The password of the manufacturer (User-3) is dynamic. If necessary,
please contact us for information.
5.5 F parameter setting and lookup table
5.5.1 Step of parameter setting
(1) in the weighing display state, according to the
Introduce the method
of login (skip this step if you have already logged in).
(2) Press and hold the [Fxx] key for about 2 seconds, and the lower row of the
instrument will display [F01] and the upper row will display [D, * * *]. You
can select different parameters by pressing the [↑] [↓] key to operate, and
the lower row will display the current F parameter number in a flash.