29
Cleaning and Maintenance
Copyright 2019 Dilon Technologies, Inc. All rights reserved. Version 1.7
Copyright 2019 Dilon Technologies, Inc. All rights reserved. Version 1.7
Note:
Prevent the handle from drying (i.e., wrap/cover in moist
germicidal wipe) and clean the CoPilot VL
®
+ as soon as feasible after
use. Patient debris can harden and become difficult to remove if
not immediately cleaned. This could inhibit the subsequent cleaning
processes.
CoPilot VL
®
+ Video Handle
Three EPA-registered disinfectants that are qualified
as intermediate level disinfectants are PDI Healthcare’s
Sani-Cloth AF# wipes (EPA reg number 9480-9), Super
Sani-Cloth wipes (EPA reg number 9480-4), and Sani-
Cloth Bleach wipes (EPA reg number 9480-8).
STEP 1: Preparation for decontamination
STEP 2: Cleaning and intermediate level disinfection
Select an appropriate quaternary ammonium isopropanol
based germicidal cleaner labeled suitable for use on
healthcare equipment and capable of intermediate level
disinfection. Reference EPA-registered disinfectants:
http://www.epa.gov/oppad001/chemregindex.htm
Thoroughly wipe to wet all exposed Video Handle
surfaces and remove any visible soil.
1.
1.
•
Immediately brush with a soft-bristled brush
moistened with an enzymatic cleaner for a minimum
of 1 minute to loosen/remove excessive visible soil.
1.
Moisten a lint free cloth with purified water. (ie: DI
water, RO water, Distilled water, etc.) Wipe handle with
2.
3.