DL9110 / DL9118
Operating Instructions
- 34 -
8.5 Replacing the Fuse.
Before opening the DL9110 / DL9118 ensure that it is
disconnected from all voltages! Electric shock danger!
All replacement fuse types are specified for ratings and
size on the battery compartment cover on the DL9110 /
DL9118.
Power the unit off by selecting the Off position on the rotary switch.
Disconnect all the test leads from the unit.
Position the DL9110 / DL9118 face down and release the captive 4
screws in the battery compartment cover.
Remove the battery compartment cover.
Lift one end of the fuse out of the fuse holder with the help of a flat
bladed screwdriver.
Lift the defective fuse completely out of the fuse holder.
Insert a new fuse as described and specified by the text on the battery
compartment cover.
Ensure that the new fuse is seated and centred in the fuse holder.
Relocate the battery cover over the battery compartment and fasten in
position with the battery cover captive screws.
8.6 Service and Calibration.
To maintain the specified accuracy of the measurement results, the
instrument must be recalibrated at regular intervals by either the
manufacturer or an authorised Service Agent. We recommend a
recalibration period of one year.