Maintenance Guide 
Veterinary Dental Cassette Kit 

  TM-723-EN 

1. Veterinary Dental Cassette

Supporting multi imaging plates on a moving plate. 

Veterinary Dental Cassette

-

Imaging plate size : Each of imaging plate size is marked on the moving plate

(1) Size 0 (22x31 mm) - 2ea,

(2) Size 1 (24x40 mm) 

– 2ea,

(3) Size 2 (31x41 mm) 

– 4ea,

(4) Size 3 (27x54 mm) 

– 2ea

(5) Size 4c (48x54 mm) - 2ea,

(6) Size 4 (76x57 mm) 

– 1ea

Veterinary Dental Cassette 

Moving Plate 

2. Imaging Plate Cleaning

Protect your investment

Dust  and  dirt  are  piled  up  on  CR  imaging  plate,  leading  to  image  artifacts.  To  avoid 

contamination, plates need cleaning at regular intervals. 

W

ARNING

 

As CR imaging plates are highly sensitive to water based cleaning 

fluids, we strongly recommend an ethanol-based plate cleaner. 

If possible, use a cleaner that is specially designed for CR imaging 

plates. This will help to ensure the longest possible service life. 

Summary of Contents for Fire CR

Page 1: ...not necessarily represent standards or specifications which must be met by Digiray Corp All information contained herein is intended for guidance purposes only and characteristics of the products and...

Page 2: ...Cassette Kit TM 723 EN 2 Contact Digiray Corp 825 Unitech Venture Town 142 Ilsan ro Ilsandong gu Goyang si Gyeonggi do 10442 Republic of Korea Tel 82 31 932 8401 Fax 82 31 927 9988 Website www digira...

Page 3: ...ges have been made to delivered system and accessories the warranty by Digiray Corp becomes void Digiray Corp will not accept any responsibility or liability for the proper functioning of the product...

Page 4: ...nd operation of this device should only be undertaken by qualified trained personnel The operator should study instructions and precautions carefully before starting to use the device listed here and...

Page 5: ...ccur unplug the device from the electrical outlet and contact authorized service personnel The power cord or power adapter is damaged An object has fallen into the device The device has been exposed t...

Page 6: ...e 4c 48x54 mm 2ea 6 Size 4 76x57 mm 1ea Veterinary Dental Cassette Moving Plate 2 Imaging Plate Cleaning Protect your investment Dust and dirt are piled up on CR imaging plate leading to image artifac...

Page 7: ...e surface of the protective layer and they appear on the image The phosphor layer has peeled away from the support layer 2 2 Imaging Plate Cleaning Tools Moving plate extracting tool Imaging plate cle...

Page 8: ...tte the Cassette Lock Releaser should be linked up with the Cassette The the Cassette Lock Releaser are pressing those buttons during it is linked up with the Cassette correctly Stand the Cassette on...

Page 9: ...hanol directly on to the imaging plate Make sure the imaging plate surface is completely dry before using the imaging plate WARNING Be sure not to damage the imaging plate Do not scratch the imaging p...

Page 10: ...ck Releaser Make sure both of buttons are pressed Insert moving plate into the cassette WARNING Do not put the imaging plate upside down when it is inserted Do not insert the moving plate in wrong dir...

Page 11: ...Kit TM 723 EN 11 DANGER Never press push buttons of the cassette with upside down position A crash will cause the damage of the imaging plate and moving plate which can affect X ray image quality Alw...