DS-425 Service Manual --- V1.04
e.g.) With Header code and weight stability code
Net weight =-1,2340 Tare weight = 0,450 and weight is stable
Null
CR
0
-
0
1
,
2
3
0
CR
4
0
0
0
,
4
5
0
CR
LF
Note 1): The data format consists of (header code), minus/plus sign, numeric data and
decimal point. If weight data is minus, -(2DH) is added after header data. If
weight is positive data, 0(30H) is output.
Note2): The data is aligned from right side.
Note3): When weight is over the capacity, the following data is output:
0OV00000CR4???????CRLF (With Header Code)
Note4): By specification setting, the decimal point can be set to period (2EH) or comma
(2CH).
4.4. Communication Method
Communication method can be set as Stream Output, Manual and Command.
4.4.1. Stream Output (Continuously Output)
Data will output to PC side continuously. Please also define the contents of data on
SPEC 03 bit 0&1) is not effective.
4.4.2. Manual Command
Data will output when
↑
key pressed, user may select to output data right away or
hold the command until weight stable by specification setting. If weight is not stable
within certain interval, Error message “ERR C” will appear on the display.
PC Side DS-425 Side
No
X. Seconds
Yes
Text
No
Weight
Stable?
Manual Command
Yes
Time Out
04 RS-232C Protocol Page No. 17