Short Form Calibration
D3655-US Rev E
EZ400 User’s Manual
19
Enter A New Setup And
Calibration Number
1. Press and hold then press
for 3 seconds to enter short
form calibration.
2. The display will flash “
SETUP
” and
then display the 6-digit setup
number with the right digit
flashing.
3. Press several times to
increment digit to it proper value.
4. Press to advance digit left.
Repeat steps 3 and 4 for each digit as
required.
5. Press to enter new setup
number and display calibration
number.
6. Repeat steps 3 and 4 to modify
the calibration number.
7. Press to enter new
calibration number and display will
go back to normal.
8. Verify the accuracy of scale.