Troubleshooting  

D3709 Rev B 

EZ2400 User’s Manual 

23 

 

FLOW CHART 

 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 

Continued 

1. Disconnect all the Load Cell 

wires from the terminal blocks 

inside the J-Box (leave the 

Indicator on while connecting 

and disconnecting the wires, it 

will not damage Load Cells or 

Indicator if wires are shorted 

during this step). Is reading on 

Indicator stable? 

2. Zero balance the Indicator. 

(Press “NET/GROSS” then 

“ZERO”). Indicator should display 

“0”. 

YES 

Replace J-Box  

(be aware of electrical 

interference that might affect 

your scale such as: mobile 

phones, CB radios, radio 

towers, electric motors, etc.). 

NO 

Note:

 Hook up the Load Cells to the J-Box one at 

a time (only one Load Cell connected at a time). 

This will get a reading for each Load Cell. While 

performing this test, watch for any other 

symptoms such as erratic/unstable display. 

Indicator flashing “±RANGE”, negative reading, 

etc. If the Indicator reading should ever appear 

abnormal with any Load Cell connected then it is 

probably bad. 

4. Record the Indicator reading with 

the Load Cell connected. 

3. Connect one Load Cell back into one of the 

terminals in the J-Box. (The reading you get for 

each Load Cell is dependent on the size and type 

of each Load Cell and how much weight is over 

each Load Cell. In general, the number should be 

positive and stable.) 

5. Stand or hang your weight over the connected 

Load Cell. Record how much the weight 

increased with your weight over the Load Cell. (A 

scale with only one Load Cell will weigh heavy.) 

Note:

 If the scale responded to your weight, 

that’s verification on the J-Box is OK. If the 

scale did not respond, either that Load Cell is 

bad or the J-Box is bad. Try the other Load 

Cells. If the Indicator still shows no response, 

the J-Box is bad. (Replace J-Box) 

6. Disconnect the first Load Cell and reconnect 

a second one. Record the Indicator reading. 

Stand or hang your weight over the connected 

Load Cell. Record how much the weight 

increased. 

7. Repeat step 6 for the remaining Load Cells. 

Remember to record your readings. 

Do not expect the Load Cells to give the same 

reading. It is common for Load Cells to have 

readings that vary by hundreds, even thousands. 

Especially when one is carrying more weight. 

8. Bad Load Cells will have a reading that is 

either unstable, makes the indicator flash 

“±RANGE” or is more than three times greater 

or less than the average of the others. Also the 

readings of your weight over each Load Cell 

should be similar. (Probably 4 times your actual 

weight). Any differences could be an indication 

of a bad Load Cell or a structural problem. 

Summary of Contents for EZ2400

Page 1: ...D3709 US June 14 2010 EZ2400 Operators Manual Ft Atkinson Wisconsin USA Panningen The Netherlands www digi star com HELLO...

Page 2: ...EZ2400 User s Manual D3709 Rev B...

Page 3: ...ons 18 Connect Load Cells to J Box 19 Load Cell Direction 19 OPTIONAL EQUIPMENT 20 Cab Control Wireless 20 Data Transfer Options 20 Transmitter Receiver 20 Remote Indicators 21 TROUBLESHOOTING 22 LEGA...

Page 4: ...Table of Contents EZ2400 User s Manual D3709 Rev B...

Page 5: ...D C 160 mA nominal with four 350 L C SET UP AND CALIBRATION Via front panel GROSS RANGE 999 999 max display LOW BATTERY WARNING Enabled at 10 5V nominal POUND KILOGRAM Selectable DISPLAY STD EZ 6 Digi...

Page 6: ...aners hoses to clean the indicator Charging Battery and Welding Disconnect all cables from the weighing indicator before charging the battery or welding on the machine If cables are left connected the...

Page 7: ...ase 3 turns indicator on Press while on runs self test 4 turns indicator off 5 temporary zero Net Mode 6 records to memory or prints displayed weight 7 Display Window Displays current actions 8 toggle...

Page 8: ...l Printer Port optional to communicate with computer and other digital input output devices 12 Remote Port optional for remote display 13 Load Cell Port for J Box cord 14 Power Port for power cord 15...

Page 9: ...t Gross Tare is a temporary zero Net Weight to display total weight Gross Weight Press 1 Press 1 Press and hold for 3 seconds to zero balance indicator 2 Flashing arrow points to gross next to the dis...

Page 10: ...d flashing arrow on side of display points to NET 3 Add more weight 4 To know total of original weight of 4000 pounds plus added 300 pounds press to show 4300 pounds flashing arrow points GROSS 5 Pres...

Page 11: ...rinting Timer Option Stopwatch for mixing time 1 Press Indicator sends data to printer or PC 1 Repeatedly press until TIMER is displayed 1 Press to displays hours minutes and seconds hh mm ss 1 10JA08...

Page 12: ...lashing digit left 2 Repeatedly press changes number 3 Press to start timer 4 When timer reaches zero press to clear timer 1 Add weight on scale Example 500 pounds 2 Repeatedly press until M is displa...

Page 13: ...nds 5 Repeatedly press until M is displayed 6 Press indicator add 1000 pounds to 500 pounds in memory and RM flashes 7 Repeatedly press until RM is displayed 8 Press 9 Total of both weights 1500 pound...

Page 14: ...m Memory Indicator must have optional printer port for printing Weight Averaging 1 Repeatedly press until RM displayed 2 Press shows weight in memory Example 1500 pounds 3 Press while weight displayed...

Page 15: ...ghts to average is two Example weight of 1000 pounds and 500 pounds averaged 4 Displays AVERAG 5 Display shows average of two weights in memory After displaying average weight indicator returns to gro...

Page 16: ...egal for trade systems 1 Repeatedly press until DIMMER is displayed 2 Press dims backlight 1 Repeatedly press gives following options Timer Stopwatch for mixing time M Adding weigh to weight memory RM...

Page 17: ...FEATURES IN MOST INDICATORS LANGUAGE langag 101 English Dutch French German Italian Portuguese Spanish Danish Hungarian Spanish ENGLSH NEDERL FRANCS DEUTSH ITAL PORT ESPAN DANSK MAGYAR VESTA Select l...

Page 18: ...m yy 6 dd mm yyyy 7 ddmoyy 8 ddmoyyyy Select date format DATE date 204 Enter XXXXXX Select key changes date function key chooses mm dd yy TARE AUTO PRINT tareap 211 ON OFF If ON tare auto prints displ...

Page 19: ...GHT HIGHWT 242 Enter analog weight value to equal 20mA or 5 volts ANALOG SELECT ANAOUT 243 Select 0 5V 4 20ma or 0 20ma output MENU 3 SCALE CALIBRATION SETTINGS DISPLAY UNIT lb kg 301 01 02 05 1 2 5 1...

Page 20: ...Numbers 1 Repeatedly press until SETUP is displayed 2 Press 3 Six digit setup number displayed Change number press moves flashing position left Press changes number 4 Press stores setup number Indica...

Page 21: ...403780 SCR 10 X 5 8 FHSTS BLACK ZP C 840459 SUPPORT HAT BRACKET D 405069 U BOLT 1 4 20 X 3 25 ZP E 405084 NUT 1 4 20 TOP LOCKING FLANGE F 403770 BRACKET WING MOUNT G 405124 PACK WEDGE MOUNT BRACKET W...

Page 22: ...Cable Connections Scale Indicator Remote Indicator Optional See J Box Connections Power Cord J Box Connection Digital Input Output Connection Optional Remote Indicator Connection Optional Power Cord...

Page 23: ...of arrow when installing load cell Connect load cell wires to terminal blocks See Wire Color Key J Box Illustrated for 4 Load Cell Installation Tighten Nuts Load Cell Cable J Box Cable Connect to Indi...

Page 24: ...mproves loading accuracy Functions Communicates with multiple mixers Specification Internally mounted 2 4 GHz radios Up to 1000 foot range 24 channels 12 or 24 volt DC system Kit Data Down Loader Allo...

Page 25: ...mall remote display RD2400V backlit remote display with 1 7 high numbers RD2400V backlit remote display w transmitter and installed receiver RD2400 backlit remote display with 1 high numbers RD2400 ba...

Page 26: ...negative side When using a multimeter to check for voltage measure between pin 1 pos and pin 2 neg Meter should read between 10 5 and 14 5 volts DC if using a tractor power cord black wire is positive...

Page 27: ...of each Load Cell and how much weight is over each Load Cell In general the number should be positive and stable 5 Stand or hang your weight over the connected Load Cell Record how much the weight inc...

Page 28: ...eatures Digi Star Legal for Trade indicators are modified to comply with US and Canadian regulatory agencies Some of these modifications are Load Cell Disconnect Sensor Indicator displays RANGE or RAN...

Page 29: ...memory and do not require batteries to maintain the audit trail values Changing any of the following parameters will change the Calibration event count value or parameter audit event count value This...

Reviews: