17
Technical specifications
Technical characteristics
• Thermal imager
Accuracy calculated as [%r (num. digits*resolution)] at 18°C ÷ 28°C <75%RH.
• DC Voltage
• AC TRMS Voltage
(*) Accuracy specified from 10% to 100% of the measuring range, sine wave. Input impedance: >9M
Ω
;
Accuracy PEAK function: ±10%rdg, PEAK response time: 1ms.
• AC+ DC TRMS Voltage
• DC Current
Field of view (FOV) / Minimum focus distance 21° x 21°/ 0.5m
Spatial resolution (IFOV)
4.53mrad
IR resolution
80 × 80 pixels
Thermal sensitivity / NETD
< 0.1°C @ +30°C (+86°F) / 100 mK
Image frequency
50 Hz
Focus mode
Focus free
Focal length
7.5mm
Focal Plane Array (FPA) / Spectral range
Uncooled microbolometer / 8–14 μm
Object temperature range
–20°C to +260°C (–4°F to +500°F)
Accuracy
±3°C (±5.4°F) or ±3% of reading (Environment temperature
10°-35°C, object temperature >0°C)
Range
Resolution
Accuracy(*)
Protection against overcharge
(50 Hz÷60 Hz)
(61 Hz÷1k Hz)
6.000 V
0.001 V
± (0.8%
reading
+ 5 digits)
± (2.4%
reading
+ 5 digits)
1000 VDC/ACrms
60.00 V
0.01 V
600.0 V
0.1 V
1000 V
1 V
Range
Resolution
Accuracy
Input impedance
Protection against overcharge
600.0 mV
0.1 mV
±(0.09%
reading
+ 5 digits)
>10M
Ω
1000 VDC/ACrms
6.000 V
0.001 V
60.00 V
0.01 V
600.0 V
0.1 V
±(0.2%reading
+ 5 digits)
1000 V
1 V
Range
Resolution
Accuracy (50 Hz÷1k Hz)
Input impedance
Protection against overcharge
6.000 V
0.001 V
±(2.4% reading
+20 digits)
>10M
Ω
1000 VDC/ACrms
60.00 V
0.01 V
600.0 V
0.1 V
1000 V
1 V
Range
Resolution
Accuracy
Protection against overcharge
600.0 μA
0.1 μA
±(0.9% r 5 digits)
Quick fuse 800mA/1000 V
6000 μA
1 μA
60.00mA
0.01mA
600.0mA
0.1mA
±(0.9% r 8 digits)
10.00A
0.01A
±(1.5% r 8 digits)
Quick fuse 10A/1000 V