10
Capturing Minimum and Maximum Values
The MAX MIN Record mode captures minimum and maximum input
values. When the input goes below the recorded minimum value or above
the recorded maximum value, the meter beeps and records the new value.
This mode is for capturing intermittent readings, recording minimum and
maximum readings unattended, or recording readings while equipment
operation precludes watching the meter.
To activate the MAX MIN mode, press soft key labeled
3
. If the meter is
already in MAX MIN function, pressing
3
causes the Meter to turn off
MAX MIN function.
Relative Values
To activate the Relative mode, press the soft key labeled
s
.
If the meter is already in the relative function, pressing
s
causes the
meter to turn off Relative mode.
Capturing Peak Values
To activate the Peak mode, press the soft key labeled
4
.
If the meter is already in the peak function, pressing
4
causes the meter
to turn off Peak mode.
Non-Contact AC Voltage Detector (100 to 1000V AC)
WARNING:
Risk of Electrocution. Before use, always test the Voltage
Detector on a known live circuit to verify proper operation.
WARNING:
Insulation type and thickness, distance from the source, and
other factors may effect operation. Always verify live voltage using other
methods before working on electrical circuits.
• The non-contact voltage detector operates when the meter is set to
any measuring function. The detector does not operate when Auto
Power Off turns the meter off or when the rotary function switch is
set to the Off position.
• Slowly move the detector probe closer to the conductor being tested.
• If AC voltage within the specified range is present, the indicator light
will illuminate.
NOTE:
The detector is designed with high sensitivity. Static electricity
and other sources of electrical energy may randomly activate the detec-
tor. This is normal operation. The detector only activates the indicator
light when AC voltage is present. It does not indicate the voltage level
on the LCD.