15
Meg OHMS
Terminal
voltage
Range
Resolution Accuracy
Test
current
Short
circuit
current
125 V
(0% to +10%)
0.125 to 4.000 M
Ω
0.001 M
Ω
±(2% +10)
1 mA @
load
125 k
Ω
≤
1 mA
4.001 to 40.00 M
Ω
0.01 M
Ω
±(2% +10)
40.01 to 400.0 M
Ω
0.1 M
Ω
±(4% +5)
400.1 to 4000 M
Ω
1 M
Ω
±(5% +5)
250 V
(0% to +10%)
0.250 to 4.000 M
Ω
0.00 1M
Ω
±(2% +10)
1 mA @
load
250 k
Ω
≤
1 mA
4.001 to 40.00 M
Ω
0.0 1M
Ω
±(2% +10)
40.01 to 400.0 M
Ω
0.1 M
Ω
±(3% +5)
400.1 to 4000 M
Ω
1 M
Ω
±(4% +5)
500 V
(0% to +10%)
0.500 to 4.000 M
Ω
0.001 M
Ω
±(2% +10)
1 mA @
load
500 k
Ω
≤
1 mA
4.001 to 40.00 M
Ω
0.01 M
Ω
±(2% +10)
40.01 to 400.0 M
Ω
0.1 M
Ω
±(2% +5)
400.1 to 4000 M
Ω
1 M
Ω
±(4% +5)
1000 V
(0% to +10%)
1.000 to 4.000 M
Ω
0.001 M
Ω
±(3% +10)
1 mA @
load
1 M
Ω
≤
1 mA
4.001 to 40.00 M
Ω
0.01 M
Ω
±(2% +10)
40.01 to 400.0 M
Ω
0.1 M
Ω
±(2% +5)
400.1 to 4000 M
Ω
1 M
Ω
±(4% +5)
1065DGMAN_20250-60 DS Insulation tester manual.indd 15