Setup and Operation

1.   Press 

Power on/off

 button to turn the unit on. The value immediately  

shown is the number of data logging sample points remaining (from  

10,000 to 0). After one second, the display will then switch to the  

temperature measurement reading mode. 

2.   Press 

MAX/MIN

 button to step through the maximum, minimum, and  

average readings. To exit the Max/Min/Avg mode, press and hold  

MAX/MIN

 button for three seconds to resume normal operation. 

3.   Press 

°C/°F/K

 button to switch between Celsius (°C), Fahrenheit (°F), and  

Kelvin (K).

4.   Press 

HOLD

 button to freeze or unfreeze the displayed readings. The screen 

shows the 

HOLD

 icon.

5.   Press 

DIF/T1–T2–T3–T4

 button to toggle between showing the T1, T2, IR,  

T1–T2, T1–IR, and T2–IR (differential temperature measurement) in the  

primary, secondary, and tertiary display. Press and hold the 

DIF/T1–T2–T3–T4

 

button for two seconds to toggle between showing T3, T4, IR, T3–T4, IR–T3, 

IR–T4 in the primary, secondary, and tertiary display. Press the 

DIF/T1–T2–

T3–T4

 button again for two seconds to exit.

6.   Press 

Setup/Backlight

 button to turn on the backlight. Press it again to turn 

off the backlight. Press 

Setup/Backlight

 button and hold for three seconds to 

enter or exit the Setup mode. (See "Changing Setup Options")

7.   Press 

 or 

Rec

/

 button to scroll to the Setup option you want to change. 

Press 

 button to increase the displayed setting.

8.   Press 

Rec

/

 button to start recording and press again to stop recording. If 

entering a Setup option, scroll to the Setup option you want to change. Press 

Rec

/

 button to decrease the displayed setting.

9.   Press 

ENTER

 button to enter a Setup option. Press 

ENTER

 button again to 

store the new setting into memory.

Using Thermocouple Input(s)

1.  Plug the type K thermocouple(s) into the T1, T2, T3, and T4  

input connector(s).

2.  Turn meter on. After one second, the thermometer displays the 

first reading. If no thermocouple is plugged into the selected  

input or if the thermocouple is "open," the display will show "- - - -".

Using Infrared Input

1.  Purchase optional infrared wand model number 20250-24  

(shown at right).

2.  Plug the IR wand into the IR input connector on side of meter;  

“IR - - - -” will show on screen. If the IR wand is not connected,  

“IR - - - -” is not shown on the screen.

3.  Press the 

SCAN

 button on the IR wand to begin measuring and 

the IR temperature is shown on the screen.

Summary of Contents for 20250-20

Page 1: ...anual THE STANDARD IN PRECISION MEASUREMENT Professional Thermocouple Thermometer with Four Type K Inputs Data Logging and Optional Infrared Wand with NIST Traceable Calibration Model 20250 20 Enter R...

Page 2: ...contact your instrument supplier immediately 1 Meter 2 Four type K flexible probes 3 USB cable 4 Software 5 Carrying case 6 One 9 V battery 7 User manual 8 NIST traceable calibration report with data...

Page 3: ...4 2 LCD 3 Enter button 4 Up button 5 Rec button 6 C F K button 7 HOLD button 8 DIF T1 T2 T3 T4 button 9 Power on off button 10 MAX MIN button 11 Setup Backlight button 12 IR input Optional infrared wa...

Page 4: ...rared IR TEMP icon requires Optional IR wand sku number 20250 24 7 SETUP mode icon 8 Offset icon 9 Auto power off icon 10 Low battery icon 11 T2 or T4 temperature display 12 MAX MIN AVG icons 13 Secon...

Page 5: ...ton to turn on the backlight Press it again to turn off the backlight Press Setup Backlight button and hold for three seconds to enter or exit the Setup mode See Changing Setup Options 7 Press or Rec...

Page 6: ...t to change this setting 3 Press or Rec button until the setting you want to use appears on the display 4 Press Enter button to store the new setting into memory Note Setup is disabled in the Max Min...

Page 7: ...f no adjust ment is needed press Enter button to exit time setting 2 Set the time The adjust order is year month day hour minute Press C F K button until the value you wish to change is flashing 3 Pre...

Page 8: ...powered off Displaying Temperatures 1 Press C F K button to switch between Celsius C Fahrenheit F and Kelvin K temperature scales 2 Hold or attach the thermocouple s to the measurement location The te...

Page 9: ...with 0 sample points remaining the first value shown will be FULL rather than a number Or if you try to record the REC icon will flash five times and disappear At this point you are no longer able to...

Page 10: ...ta List tab to open the list of measurements recorded Select File icon from the menu bar and save data to desired location on your PC Note Data will be saved in an Excel format that will include all s...

Page 11: ...bove 14 F 10 C 11 F 6 C below 14 F 10 C Auto power off Unit shuts off automatically after 20 minutes to conserve battery life Operating temperature 32 to 122 F 0 to 50 C Storage temperature 14 to 122...

Page 12: ...ery Replacement When the battery power falls low the low battery icon will appear on the screen Replace the 9 V battery in the rear battery compartment by remov ing the cover Ensure that the cover is...

Reviews: