Alarm Relays (TH625)

The SPST 24V 500mA relay contacts are normally open and will close on alarm conditions when the alarms 
are enabled.  Relay contacts are always functional when the alarm is enabled.  The relay will close only 
during minimum and maximum alarm conditions.  

Dip Switch Set-Up

To set-up the TH6 recorder for your specific 

application, you might need to change some of the 

Dip Switches. The Dip Switches are located on the 

back of the unit (figure 3). Use a pen or small screw 
driver to flip the switches. 

Remember to install 

correct chart to match corresponding 

1

2

3

4

5

6

7

8

Slide toggle up for on position.

Slide toggle down for off position.

Dip Switch 1, 
7 & 8 are not 

active on this 

recorder.

Temperature Range

The TH6 will record up to 4 temperature ranges  
(depended on model). Dip switch #3 & #4 allow 
you to select the temperature range.

0 to +100°F  
(TH623/25)

#3 down 
#4 down

+32 to +100°F  
(TH621/22)

#3 down 
#4 down

0 to +185°F 
(TH623/25) 

#3 down 

#4 up

0 to +50°C 
(TH621/22/23/25) 

#3 up 

#4 down

0 to +85°C 
(TH623/25)

#3 up 
#4 up

Dip Switches for ranges

Recording Time

The TH6 recorder has two different recording time 
options

24 hour

#2 up

7 day

#2 down

NOTE: Remember to install correct chart to match 
corresponding switch setting.

3

4

D

D

D

D

D

U

U

D

U

U

DICKSON

 TH6 Dip Switch Settings: 

Up = On; Down = Off

Dickson

  -  Addison, Il    www.dicksondata.com

Remember to install the correct chart for your range

7 Day

Switches 2-4

For Model

Chart

2

D

U

All Models

All Models

TH623/25

TH621/22

0 to 100

0 to 200

0 to 100

TH623/25

All Models

TH623/25

24 Hour

0 to +100°F

+32 to +100°F

0 to +185°F

0 to +50°C

0 to +85°C

#25201

Summary of Contents for TH6

Page 1: ...fications Applications Quick Start Operating Instructions Charts Accessories Calibrations Troubleshooting Warranty Spanish French German Italian DICKSON TH6 Temperature Chart Recorder Contents Product...

Page 2: ...ameter 6 Battery Life Avg 72 hours Backup Only Recording Times 24 Hour 7 Day Response Time 2 5 minutes Alarm TH623 25 Audio Visual on High Low Temp RH Limits Relay Type SPST 24VDC 0 5A NO NC Relay doe...

Page 3: ...just the chart time press and hold ADJUST UP and ADJUST DOWN buttons located on the back of the unit next to the dip switch see figure 3 The green LED will blink rapidly for about five seconds then th...

Page 4: ...larm min e pen alarm max In order to set pen alarm minimums and maximums pressing ADJUST UP will increase the alarm value and pressing ADJUST DOWN will decrease the alarm value There is acceleration i...

Page 5: ...or on position Slide toggle down for off position Dip Switch 1 7 8 are not active on this recorder Temperature Range The TH6 will record up to 4 temperature ranges depended on model Dip switch 3 4 all...

Page 6: ...e Adjust Up button the led will then blink amber at which point only the pen being adjusted will show on the display To raise the unit of measurement press the Adjust Down button To lower the unit of...

Page 7: ...dirt dust and normal handling can throw your precision instrumentation out of calibration Regular calibrations ensure that you receive the most accurate readings possible How often should I recalibrat...

Page 8: ...y instrument The model number serial number and a purchase order number will be requested before an RA number is issued Carefully repack the instrument label the outside of the box with the RA and ret...

Page 9: ...de la punta del estilete en el cuadrante b Esta caracter stica s lo debe utilizarse para ajustes finos Para ajustar el tiempo del gr fico pulse y mantenga apretados los botones ADJUST UP y ADJUST DOWN...

Page 10: ...UST UP para aumentar el valor de la alarma y ADJUST DOWN para disminuir su valor Se produce aceleraci n si se mantiene apretado el bot n ADJUST UP Pulse repetidamente el bot n PEN HOME para que se des...

Page 11: ...Italian 1 2 3 4 5 6 7 8 Conmutaci n deslizante hacia arriba para posici n de conexi n Conmutaci n deslizante hacia abajo para posici n de desconexi n Rangos de Temperatura El TH6 Puede registrar hasta...

Page 12: ...r b Cette fonction ne devrait tre utilis e que pour l affinage de l ajustage Pour ajuster l heure du graphe pressez et maintenez enfonc s les boutons ADJUST UP et ADJUST DOWN situ s l arri re de l uni...

Page 13: ...OWN pour la diminuer Il y aura une acc l ration si le bouton ADJUST UP est maintenu enfonc Une pression r p t e de PEN HOME fera d filer les cinq options jusqu ce que soit press le bouton d alarme our...

Page 14: ...ctions Charts Accessories Calibrations Troubleshooting Warranty Spanish French German Italian Port e de temp rature L appareil TH6 enregistrera jusqu cinq port es de temp rature selon le mod le Le com...

Page 15: ...e der W hlerscheibe b Dieses Feature sollte nur zur Feineinstellung verwendet werden Um die Diagrammzeit einzustellen passen Sie die Diagrammzeit an dr cken und halten Sie die neben dem Abblendschalte...

Page 16: ...den Stift wird durch Dr cken von ADJUST UP der Alarmwert erh ht und durch Dr cken von ADJUST DOWN der Alarmwert erniedrigt Durch Gedr ckthalten des ADJUST UP Buttons wird dieser Einstellvorgang besch...

Page 17: ...ts Accessories Calibrations Troubleshooting Warranty Spanish French German Italian Temperaturbereich Der TH6 zeichnet bis zu f nf Temperaturbereiche auf in Abh ngigkeit vom Modell Abblendschalter Nr 3...

Page 18: ...ul quadrante b Questa funzione deve essere usata solo per regolazioni di precisione Per regolare l ora del diagramma premere e tenere premuti i pulsanti ADJUST UP e ADJUST DOWN che si trovano a tergo...

Page 19: ...rme e premendo ADJUST DOWN si diminuisce il valore dell allarme Si accelera se il pulsante ADJUST UP tenuto premuto Premendo ripetutamente il pulsante PEN HOME si scorrono le 5 possibilit fino a quand...

Page 20: ...s Charts Accessories Calibrations Troubleshooting Warranty Spanish French German Italian Campo temperatura Il TH6 registrer fino a 4 campi di temperatura a seconda del modello Gli interruttori Dip 3 4...

Reviews: