Dickson KT6P Manual V3 

 Why does the display read PROB?  

 

Check the sensor pod and KTC probe(s).  Make sure it is properly plugged into the unit. 

 

Is the unit locked up? This can be confirmed by pressing any of the buttons on the 
keypad, if the unit is locked up there will be no response to button presses and the unit 
may appear to be working, but the readings won't change, also the chart will not rotate. 

Remove power and battery, and then re-power. 

 Why does calibration seem to be off?  

 

What is the tolerance of the unit it's being compared to?  

 

It's ok if the unit is within the sum of the two tolerances.  

 

Has an outside calibration house attempted calibration? It may not have been adjusted 

properly.  

 

Walk through calibration adjustment, found in the manual 

 Why won't the battery back-up work?  

 

Are there good batteries in the Chart Recorder? 

 

Keep in mind that battery backup varies significantly depending on temperature, pen 

movement and chart rotation speed.  

 

If there was a momentary power outage (Brown out), the unit may not have had enough 
time to recognize this and switch to battery mode. (The Chart Recorder may lock up or 

shut off in this situation) This situation may also occur if the unit is plugged into an outlet 
that shares the circuit with other machinery that has phase motors or compressors that 
cycle periodically. As these other machines cycle, they momentarily have a high current 
draw, therefore drawing from the unit. 

  Why won't the chart recorder respond to changes?  

 

Is the unit locked up?  This can be confirmed by pressing any of the buttons on the 

keypad, if the unit is locked up there will be no response to button presses and the unit 
may appear to be working, but the readings won't change, also the chart will not rotate. 
Remove power and batteries, and then re-power.  
 

  Why won't the unit power up?  

 

Remove the batteries and power adapter for a minute or two, this will reset the unit. The 
unit should respond when the adapter is plugged in. 

 

WARRANTY 
 

Dickson warrants that this line of instruments will be free from defects in material and 
workmanship under normal use and service for a period of twelve months after delivery. 
 

This warranty does not cover routine calibration and battery replacement. 

 

 

 
 
 

 
 

www.

.com

[email protected]

1.800.561.8187

Summary of Contents for KT6P1

Page 1: ...he timing arrow just to the right of the pen tip on the dial Press Pen Home to move the pen s back onto the chart The following feature should be used for fine adjustments only b To adjust the chart t...

Page 2: ...set the alarm make sure unit is on and press and hold the Alarm button The LED will blink red rapidly for about five seconds then the LED will turn solid green Release the Alarm button and the LED wi...

Page 3: ...ill close on alarm conditions when the alarms are enabled Relay contacts are always functional when the alarm is enabled The relay will close only during minimum and maximum alarm conditions Relays wi...

Page 4: ...d green Release the Pen Home and On Off buttons The LED will flash amber and green for one second then the LED will turn off Note Press Pen Home first so the unit will not turn off 2 The pen will move...

Page 5: ...tion If you wish to cancel the Calibration simply enter Calibration mode and toggle through the steps without adjusting displayed readings Exit by pressing the On Off button You have now restored fact...

Page 6: ...d enough time to recognize this and switch to battery mode The Chart Recorder may lock up or shut off in this situation This situation may also occur if the unit is plugged into an outlet that shares...

Page 7: ...t to operate properly Calibration Only the Replaceable Sensor is calibrated The sensor and all calibration defaults and adjustments are stored on the Replaceable Sensor Accuracy is strictly controlled...

Reviews: