OPERATOR’S MANUAL
mini GAC
®
11001-1464A-201701 Rev B
QUICK START GUIDE / 5
QUICK START GUIDE
IMPORTANT: The cell must be empty and the loading cup
removed before turning the unit ON and
during all testing (progress bar indicates
when testing).
SELECTING GRAIN
1. Press the
On/Off/Home
button to power on the unit.
2. To select a grain type, press the
Down Arrow
button to
position cursor at the
Product
heading and press
Enter
.
3. Select a product (grain) using the
Up or Down Arrow
button
and press
Enter
.
4. The selected grain will appear at the top of the display.
GRAIN SAMPLE TESTING
1. Ensure Loader is removed from the top of the unit.
2. With the product name selected at the top of the display,
press the
Enter
button.
3. The cell must be empty, upright, and still to perform an
empty cell test. When the Empty Cell menu displays, press
the
Enter
button.
IMPORTANT: The empty cell reference measurement is
used for all product testing until the unit is
powered off. It is critical that the cell be
clean and completely empty when the unit is
on. Keep the measurement cell opening
clear of hands or other objects during this
period.
4. After the empty cell test completes, the
Fill Cell
menu
displays.
5. With the slide closed, scoop or pour grain into the Loader
over the minimum fill line (Label on loader identifies
minimum fill line) (
6. Place the Loader on top of the unit.
7. When secured, pull Loader slide out to dispense grain into
the unit.
8. Remove the Loader and use the level edge to strike
any excess grain from the unit.
9. Press
Enter
to begin the test.
NOTE: Refer to the Normal
Operation section for
additional information on
the screen displays of
Selecting and Testing
Grain.