OPERATOR’S MANUAL
GAC® 2500 AGRI
11001-1626B-201405
SETUP
/
41
LCD DISPLAY TOUCH SCREEN CALIBRATION
The LCD display touch screen may not respond as it should with a finger
touch or stylus after a period of time or if subjected to extreme temperature
changes. The display can be re-calibrated to improve responsiveness.
1. At the Service Data screen, press the
Display Touch Screen
Calibration
button and follow onscreen instructions.
2. Carefully press and hold finger on the center of the target.
3. Repeat as the target moves around the screen.
4. With a keyboard attached, press the
Enter
button to accept new
settings or the
Esc
button to keep previous settings. Without an
attached keyboard, press anywhere on the screen to accept.
Figure 36
Display Calibration
Carefully press and briefly hold stylus on the center of the target.
Repeat as the target moves around the screen.
Press the Esc key to cancel.
+
Press Target
with finger
or stylus
New calibration settings have been measured.
Press the Enter key to accept the new settings.
Press the Esc key to keep the old settings.
+
+
+
+
Service Data
Service Data
Service Note:
Frank from DICKEY-john 457-2222
basic checkout
Service Date:
03 10
/ /2013
Resettable Counter :
Total Cycles:
Reset
2000
100,000
GAC2500-AGRI
3
4
5
Summary of Contents for GAC 2500-AGRI
Page 4: ...OPERATOR S MANUAL GAC 2500 AGRI 11001 1626B 201405 IV...
Page 20: ...OPERATOR S MANUAL GAC 2500 AGRI 11001 1626B 201405 16 NAVIGATION...
Page 68: ...OPERATOR S MANUAL GAC 2500 AGRI 11001 1626B 201405 64 GRAIN CALIBRATIONS...
Page 70: ...OPERATOR S MANUAL GAC 2500 AGRI 11001 1626B 201405 66 DIAGNOSTICS...
Page 82: ...OPERATOR S MANUAL GAC 2500 AGRI 11001 1626B 201405 78 MAINTENANCE...
Page 84: ...OPERATOR S MANUAL GAC 2500 AGRI 11001 1626B 201405 80 TROUBLESHOOTING...