8. APPENDICES
8.1. Warning flags
Uppercase letters refer to WBC or HGB problems:
Flag
Meaning
Recommended user action
E
No WBC 3-part
differential
Possible lyse problem. May occur in pathological
lymphocytosis.
H
HGB blank is high, or no
HGB blank
Repeat blank measurement. If HGB blank is not
stable, there are probably bubbles in the WBC
chamber: Run a cleaning and try blank again. Close
the side door if it was open during measurement.
B
WBC blank is high, or no
WBC blank
Repeat blank measurement, or run prime lyse and try
blank again.
Possible lyse contamination, or noise problem.
C, Q
WBC clogging
Aperture clogging. Perform cleaning and repeat the
measurement. If it is a general problem, please
contact your Service Personnel.
Low temperature reagents can cause it as well (mainly
diluent), in this case you will have to wait until they
reach room temperature.
Warning flags in lowercase refer to RBC or PLT problems:
Flag
Meaning
Recommended user action
p
PLT blank is high, or no
PLT blank
Run cleaning and repeat the blank measurement.
Diluent or system cleanliness problem. If it is stable high,
replace the diluent by opening a new tank.
b
RBC blank is high, or no
RBC blank
Same action as in case of warning flag
p
.
c
RBC/PLT clogging
The same action as in case of the
C
warning flag.
Result Warning Flags
Flag Meaning
Description
X
EOS Blank high
Similar to the WBC B flag. Run a cleaning cycle and
rerun blank.
Y
Clogging
The LyseEO measurement has encountered a clogging
in the aperture. Run a cleaning cycle and repeat sample
measurement
Z
Noise in the EOS Channel
Possible lysing problem. Check if there is enough
reagent in the container. Prime the LyseEO reagent.