52
6.5.2. Abacus4 message code list
Code
Message text
5001
Remove reagent tubing at rear reagent inputs (Diluent, Lyse, LyseEOS and
Cleaner). Leave waste connected.
5002
Connect min. 100 ml distilled water to reagent inputs using cleaning tube kit.
5003
Remove cleaning tube kit. Keep reagent inputs free.
5004
You can power off the system!
5005
Apertures are partially clogged. Try cleaning!
5006
Please empty waste container!
5101
%u data record(s) is selected.
5102
Insert data disk!
5106
No such type of data
5200
New reference ranges will be saved. Press OK to confirm.
5300
Some or all of the calibration factors are out of range!
5400
+12V too low!
5401
+12V too high!
5402
Battery voltage too low!
5901
Insert software update USB drive!
The system will restart!
5902
Are you sure to delete all device statistics?
5903
This function will delete all measured results.
Are you sure?
5904
This function will delete all selected records.
Are you sure?
5905
Are you sure %s?
5906
Now it is safe to turn the instrument off...
5907
Now it is time to clean the washing head with a damp cloth, and check the pump
cleanliness. Make sure they are all clean, then you can proceed.
5908
Set the needle and press OK to continue!