Code
Error text
Reason(s)
Remedy
1401
Printer error. Check paper!
Printer is out of
paper and there are
some other printer
errors!
Feed the printer with paper. Check the
connections between the instrument and
the printer and check the printer setup in
the Settings/Printer settings sub-menu.
1402
Printer is out of paper! Check
paper!
Printer is out of
paper!
Feed the printer with paper.
1403
Printer was not set up correctly The selected printer
type does not match
to the printer!
Modify the printer setup in the
Settings/Printer settings sub-menu.
1405
Paper height is too large
Paper height
overstep the
margin!
Modify the paper setup in the
Settings/Printer settings sub-menu.
1411
Error #%d in printer
initialization!
Printer software
error!
Check the connections between the
instrument and the printer and check the
printer setup in the Settings/Printer
settings sub-menu.
Try to make a software upgrade.
1900
You have to add at least one
user to use multi user mode!
Multi-user
problem.
Follow the instruction, or change to single
user mode in the User settings submenu.
1921
%s sensor calibration failed!
The system turned off the %s
sensor.
Reagent container
became empty, or
there are lot of
bubbles in reagent
tubes.
Replace the reagent container with a filled
one.
If this is still a problem with a filled
reagent container, turn on the sensor
and try recalibrating and check the
tubings and the fittings.