Tab. 1: Electromagnetic Emissions
Tab. 2: Electromagnetic Immunity
The
PulsePen
is suitable for use in the specified electromagnetic environment. The user of the
PulsePen
should
assure that it is used in an electromagnetic environment as described below.
Emissions tests
Compliance
Electromagnetic Environment
RF Emissions
CISPR 11
Group 1
The
PulsePen
uses RF energy only for its internal function.
Therefore, the RF emission is very low and not likely to cause any
interference in nearby electronic equipment.
RF Emissions
CISPR 11
Class B
The
PulsePen
is suitable for use in all establishments,
including domestic establishments and those directly connected to
the public low-voltage power supply network that supplies
buildings used for domestic purposes.
Harmonic Emissions
IEC 61000-3-2
Not Applicable
Voltage fluctuations /
Flicker Emissions
IEC 61000-3-3
Not Applicable
The
PulsePen
is suitable for use in the specified electromagnetic environment. The user of the
PulsePen
should
assure that it is used in an electromagnetic environment as described below.
Immunity
Test
IEC 60601-1-2
Test level
Compliance
level
Electromagnetic Environment
Electrostatic
Discharge (ESD)
IEC 61000-4-2
±8 KV contact
±15 KV air
IEC 60601-1-2
Test level
Floors should be wood, concrete or ceramic tile.
If floors are covered with synthetic material, the
relative humidity should be at least 30%.
Electrical Fast
Transient / Burst
IEC 61000-4-4
±2 KV
Not Applicable
Mains power quality should be that of a typical
commercial or hospital environment.
Surge
IEC 61000-4-5
±1 KV differential
mode
±2 KV common mode
Not Applicable
Mains power quality should be that of a typical
commercial or hospital environment.
Voltage dips, short
Interruptions
and Voltage
variations on power
supply input lines
IEC 61000-4-11
0% U
T
, 0.5 cycle
@ 0°, 45°, 90°, 135°,
180°, 225°, 270°, 315°
0% U
T
, 1 cycle /
70% U
T
, 25/30
cycles , @ 0°
0% U
T
, 250/300
cycles
Not Applicable
Mains power quality should be that of a typical
commercial or hospital environment. If the user
of the
PulsePen
requires continued operation
during power mains interruptions, it is
recommended that the computer used with the
PulsePen
be powered from an uninterruptible
power supply or a battery.
!
10
User Manual
PulsePen
- V. 4.6