4.3 I
NSTALLATION
OF
THE
REVERSE
OSMOSIS
MEMBRANE
1. Remove the 1 µm-sediment filter cartridge and the in-line resin cartridge from the two
plastic stirrups (Figure 1).
2. Disconnect the water inlet pipe from the membrane holder by disconnecting the quick
coupling (Figure 2).
3. Remove the membrane holder from the two plastic stirrups (Figure 3).
4. Unscrew (by turning to the left) the striped part on the top («big cap») of the membrane
holder.
5. Once the membrane holder is opened: insert the new membrane (Figure 4), with the
peripheral black seal on the top, until the membrane reaches the very bottom of the
membrane holder. The extremity of the collecting tube must then come up to the top of the
membrane holder (Figure 5).
6. Check that the O-ring is well-positoned at the rounded section of the membrane holder
(Figure 5).
7. Screw the top part (“big cap”) of the membrane holder.
8. Reconnect the water inlet pipe.
9. Clip the membrane holder, the 1 µm-sediment filter cartridge and the in-line resin cartridge
on its plastic stirrups
.
O-Classic
operator’s manual
version: 20.01.22
9
Figure 1
Figure 2
Figure 4
Figure 3
Figure 5