5
P
UTTING
INTO
SERVICE
1. Check all the hydraulic couplings.
2. Close the storage tank valve.
3. Open the purified water valve.
4. Open the tap water valve.
5. Plug the water purifier to power supply.
6. The controller goes by the following phases:
o
The POWER light indicates that the water
purifier is switched ON.
o
The PUMP light indicates that the pump is
working.
o
The FLUSH light indicates that the system is
rinsing the membrane.
The FLUSH light is turning OFF when the
system starts producing.
7. Wait (a few minutes) that the purified water flows from the tap and let flow around 5 liters.
The conductivity light will become green.
8. Proceed to a conductivity measurement (see paragraph 6.4.2). The measured value must be
between 0 and 0,1 µS/cm. Close the purified water valve.
9. Open the storage tank valve and wait until it is full: the FULL light is becoming green.
10. The controller goes to the next phase:
11. The water purifier is now operational.
O-Classic
operator’s manual
version: 20.01.22
10