P R O L Y T E
A N A L Y Z E R
3.s Electrode
Jack
Removal
1.
Insert the electrode jack wires into the supplied 11/32 inch nut driver.
2.
Remove the hex nuts using the nut driver.
3.
Pull the electrode jacks from the front housing of the analyzer, as shown in Figure P16.
3.t Electrode
Jack
Installation
1.
Push the electrode jacks through the holes in the front of the analyzer, see Figure P16.
2.
Insert the electrode jack wires into the 11/32 inch nut driver.
3.
Attach the hex nut using the nut driver and tighten.
37