P R O L Y T E
A N A L Y Z E R
3.q Guide
Rod
Removal
1.
Remove the sample probe from the probe arm and the solutions valve. Remove the solutions
valve.
2.
Detach the screw from the bottom of the guide rod, using a Phillips head screw driver. Remove the
washer.
3.
Remove the guide rod and probe arm assembly, down and away from the housing.
3.r
Guide Rod Installation
1.
Install the guide rod. To install, first insert the guide rod into the alignment hole located in the top
wall of the front assembly.
2.
Attach the washer and screw to the bottom of the guide rod, using a Phillips head screw driver.
35