Troubleshooting
Diablo 5000B Real-Time Gas Analyzer Getting Started Manual
39
voltage to reduce the Total Ion Signal.
Set the Electron Multip
lier Voltage mode to “Relative” (relative to the voltage in the tune file
used for acquisition) and enter a negative voltage offset. You may have to experiment to
determine the optimum offset for your application.
3) If none of these options are compatible with your application needs, then you may need to
either ratio results to an internal standard or recalibrate the system on a regular basis.