2. Introduction

Thank you for choosing to use the 

VENTURI

 MINO Negative Pressure Wound Therapy (NPWT) system from Direct 

Healthcare Group. The 

VENTURI

 MINO NPWT system is indicated for use in a wide range of patients with acute or 

chronic wounds.

NPWT is applied by the application of negative pressure to a sealed (occlusive) dressing placed over the wound area. 

VENTURI

 MINO Wound Care Sets include a choice of gauze or foam dressing, silicone portal drain and transparent 

adhesive film.

The 

VENTURI

 COMPACT NPWT system is intended to be reusable and will benefit from careful installation and use, 

providing a long and effective service life. Please read and understand this document completely before applying 
NPWT.

3. Important Information

3.1 Intended Use

The VENTURI MINO NPWT system may be used:
•  to help expedite wound closure (when healing by secondary intention)
•  to reduce wound size
•  for preparing a wound for closure by primary intention
•  in wound management (exudate management)
•  to downscale complexity of reconstructive surgery
Wounds which may benefit from the application of NPWT include pressure ulcers; dehisced surgical wounds; diabetic/
neuropathic foot ulcers; venous leg ulcers; post-op surgical wounds (including flaps and grafts); traumatic wounds; 
pre-op flap/graft; necrotising fasciitis; burns.

3.2 Intended Environments

The devices are intended for use by qualified healthcare professionals and lay persons in a healthcare or home 
environment.

3.3 Contraindications for Use

Do not place NPWT dressings directly in contact with exposed blood vessels, anastomotic sites, organs or nerves.
NPWT is contraindicated for patients with:
1. Malignancy in the wound
2. Untreated osteomyelitis
3. Non-enteric and unexplored fistulas
4. Wounds with difficult haemostasis
5. Necrotic tissue with eschar present
NB: After debridement of necrotic tissue and complete removal of eschar, NPWT can be used.

3.4 List of Components

Your 

VENTURI

 MINO NPWT system should comprise the following items - please ensure you have all of these before 

installation.
NB: Wound Care Sets are supplied separately.

•  VENTURI

 MINO (TG600/14)

•  5V Mains adapter FW7662M/05
•  150ml Canister (supplied fitted to power unit)
•  Carry bag (may not be included in all markets)

D H G - H E A LT H C A R E . C O M

3

U S E R   M A N U A L

Summary of Contents for TG600/14

Page 1: ...DHG HEALTHC AR E C O M DHG HE ALTHCAR E C O M VENTURI MINO User Manual...

Page 2: ...VENTURI MINO DHG HEALTHCARE COM...

Page 3: ...tions and Information 4 4 How to Apply NPWT 5 Getting Started 5 Operating the Vacuum Power Unit 5 5 Operation Guidelines 6 User Information 6 Operation Controls 7 Battery Information 7 6 Care and Main...

Page 4: ...n Warning Refer to instruction manual booklet Medical Devices Directive 93 42 EEC Medical Device Regulation 2017 745 North America ETL listed Class II Equipment Double Insulated Do not dispose of with...

Page 5: ...gery Wounds which may benefit from the application of NPWT include pressure ulcers dehisced surgical wounds diabetic neuropathic foot ulcers venous leg ulcers post op surgical wounds including flaps a...

Page 6: ...near to material which is flammable or can be damaged by heat The supplied power adapters plug directly into the power source partly mitigating this issue The VENTURI MINO TG600 14 vacuum power unit...

Page 7: ...o the VENTURI MINO power unit by lining up the two parts pushing together and twisting to lock Fig 1 Remove the bung from the connector port this can be retained to cap the used canister prior to disp...

Page 8: ...ubing as above this can be reconnected to new canister and unclamped if wound dressing is not being changed Rotate canister to unlock and remove from power unit Dispose of used canister according to l...

Page 9: ...begin operation at the default pressure of 80mmHg The selected vacuum level is shown on the display screen 5 3 Battery Information A fully charged battery should operate the power unit continuously fo...

Page 10: ...ocial care facilities TECcare CONTROL products provide class leading broad spectrum high level disinfection with an exceptional safety profile Being pH neutral TECcare CONTROL can be universally used...

Page 11: ...heck that low flow is not caused by pinched or bent tubing If exudate is pooling in the wound bed the dressing should be changed This alert will self cancel after 5 minutes and will re occur if the co...

Page 12: ...above mains adapter is considered part of the ME equipment The VENTURI MINO power unit must only be used with the specific external power adapter as supplied by Direct Healthcare Group EXPECTED SERVIC...

Page 13: ...ntinuous resets Try to ascertain the source of the interference by turning nearby or suspect equipment off and see if the interference effects stop In any such event the user is encouraged to try to c...

Page 14: ...UT 0 5 cycle At 0 45 90 135 180 225 270 and 315 Mains supply quality for the mains adapter should be that of a typical commercial and or hospital environment In the event of a mains interruption the V...

Page 15: ...rs such as base stations for radio cellular cordless telephones and land mobile radio AM and FM radio broadcast and TV broadcast cannot be predicted theoretically with accuracy To assess the electroma...

Page 16: ...et Swatar BKR 4013 Malta Direct Healthcare Group Ltd Withey Court Western Industrial Estate Caerphilly United Kingdom CF83 1BF T 44 0 800 043 0881 F 44 0 845 459 9832 E info directhealthcaregroup com...

Reviews: