6. Care and Maintenance

6.1 Power Unit

Always disconnect the 

VENTURI

 MINO power unit from the power adapter and the power adapter from the power 

source before carrying out cleaning. Check all electrical connections and power cord for signs of excessive wear. The 
power unit / power adapter can be wiped down with detergent or disinfectant solution or wipe*. Do not use solvents. 
Unsuitable for sterilisation. Dispose of the power unit / power adapter in accordance with the local regulations 
including WEEE requirements. The power unit / power adapter should be cleaned between patient use as a minimum.

*  In line with the MHRA Medical Device Alert (MDA/2013/019), Direct Healthcare Group advises customers to 
use pH neutral, high level disinfectant cleaning products to sanitise reusable medical devices to prevent damage 
to materials and the degradation of plastic surfaces after prolonged use. The use of inappropriate cleaning and 
detergent materials on medical equipment could damage surfaces and may compromise the ability to decontaminate 
medical devices adequately or may interfere with device function. Direct Healthcare Group recommends the use of 
TECcare

®

 CONTROL antimicrobial wipes and fluid to clean and decontaminate all products it supplies to health and 

social care facilities. TECcare CONTROL products provide class leading broad spectrum, high level disinfection with an 
exceptional safety profile. Being pH neutral TECcare CONTROL can be universally used on all hard and soft surfaces 
without any detrimental effect. TECcare CONTROL is CE marked for cleaning medical equipment.

6.2 Wound Care Sets and Canisters

Wound Care Sets and canisters are disposable and intended for single use only. After use please dispose of in an 
appropriate manner in accordance with local regulations and hospital best practice.

6.3 Servicing

The 

VENTURI

 MINO power unit is non-serviceable. Should a fault occur, please refer to the ‘Fault Finding’ section 

on page 9. If the power unit fails to operate correctly please contact Direct Healthcare Group or an authorised dealer.

6.4 Transport and Storage

Handle with care. Please report instances of damage or impact to Direct Healthcare Group Service Department.
–25 °C without relative humidity control; and
+70 °C at a relative humidity up to 93 %, non-condensing.
An atmospheric pressure range of 700 hPa to 1,060 hPa.
Suitable for all standard modes of transport when in the correct packaging.

6.5 Operational Conditions

A temperature range of +5 °C to +40 °C;
A relative humidity range of 15% to 93%, non-condensing; and
Operational Atmospheric Pressure: 700 hPa to 1,060 hPa
Suitable for pollution degree 2
Operational altitude ≤ 2 000 m
IP Rating: IP22 power unit only

D H G - H E A LT H C A R E . C O M

8

V E N T U R I

®

  M I N O

Summary of Contents for TG600/14

Page 1: ...DHG HEALTHC AR E C O M DHG HE ALTHCAR E C O M VENTURI MINO User Manual...

Page 2: ...VENTURI MINO DHG HEALTHCARE COM...

Page 3: ...tions and Information 4 4 How to Apply NPWT 5 Getting Started 5 Operating the Vacuum Power Unit 5 5 Operation Guidelines 6 User Information 6 Operation Controls 7 Battery Information 7 6 Care and Main...

Page 4: ...n Warning Refer to instruction manual booklet Medical Devices Directive 93 42 EEC Medical Device Regulation 2017 745 North America ETL listed Class II Equipment Double Insulated Do not dispose of with...

Page 5: ...gery Wounds which may benefit from the application of NPWT include pressure ulcers dehisced surgical wounds diabetic neuropathic foot ulcers venous leg ulcers post op surgical wounds including flaps a...

Page 6: ...near to material which is flammable or can be damaged by heat The supplied power adapters plug directly into the power source partly mitigating this issue The VENTURI MINO TG600 14 vacuum power unit...

Page 7: ...o the VENTURI MINO power unit by lining up the two parts pushing together and twisting to lock Fig 1 Remove the bung from the connector port this can be retained to cap the used canister prior to disp...

Page 8: ...ubing as above this can be reconnected to new canister and unclamped if wound dressing is not being changed Rotate canister to unlock and remove from power unit Dispose of used canister according to l...

Page 9: ...begin operation at the default pressure of 80mmHg The selected vacuum level is shown on the display screen 5 3 Battery Information A fully charged battery should operate the power unit continuously fo...

Page 10: ...ocial care facilities TECcare CONTROL products provide class leading broad spectrum high level disinfection with an exceptional safety profile Being pH neutral TECcare CONTROL can be universally used...

Page 11: ...heck that low flow is not caused by pinched or bent tubing If exudate is pooling in the wound bed the dressing should be changed This alert will self cancel after 5 minutes and will re occur if the co...

Page 12: ...above mains adapter is considered part of the ME equipment The VENTURI MINO power unit must only be used with the specific external power adapter as supplied by Direct Healthcare Group EXPECTED SERVIC...

Page 13: ...ntinuous resets Try to ascertain the source of the interference by turning nearby or suspect equipment off and see if the interference effects stop In any such event the user is encouraged to try to c...

Page 14: ...UT 0 5 cycle At 0 45 90 135 180 225 270 and 315 Mains supply quality for the mains adapter should be that of a typical commercial and or hospital environment In the event of a mains interruption the V...

Page 15: ...rs such as base stations for radio cellular cordless telephones and land mobile radio AM and FM radio broadcast and TV broadcast cannot be predicted theoretically with accuracy To assess the electroma...

Page 16: ...et Swatar BKR 4013 Malta Direct Healthcare Group Ltd Withey Court Western Industrial Estate Caerphilly United Kingdom CF83 1BF T 44 0 800 043 0881 F 44 0 845 459 9832 E info directhealthcaregroup com...

Reviews: